1.1.1
Operation and Maintenance
Manual
Electronic Laboratory Autoclaves
Models 2840 ELV-D
ELV-D: standard autoclave
C: optional system for fast cooling
PV: optional vacuum pump
BH: optional bio hazard filtration
WR: water recycling system
F: optional fan for super-fast cooling
Tuttnauer Europe B.V.,
Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda,
The Netherlands Tel: 31 (0) 765423510, Fax: 31 (0) 765423540
Cat. No. MAN205-0468001EN Rev H

Table of Contents

Other manuals for Tuttnauer 2840 ELV-D

Questions and Answers

  • G
    grace22Sep 8, 2025
    What does 'Low Pressure' mean on my Tuttnauer Laboratory Equipment and how to resolve it?
    • B
      Barbara LaneSep 9, 2025
      The 'Low Pressure' message on your Tuttnauer Laboratory Equipment indicates that the chamber pressure dropped below the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. Perform a new cycle.
  • M
    michaelgallowayAug 25, 2025
    What does 'Low Pressure' mean on Tuttnauer Laboratory Equipment?
    • M
      Martin SimsAug 25, 2025
      If you see 'Low Pressure' displayed on your Tuttnauer Laboratory Equipment, it means the chamber pressure dropped below the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. To address this, perform a new cycle.
  • A
    allenbethanyAug 25, 2025
    What does 'High Temp. (Ending)' mean on Tuttnauer 2840 ELV-D Laboratory Equipment?
    • C
      charlesclarkAug 26, 2025
      If your Tuttnauer Laboratory Equipment displays 'High Temp. (Ending)', it means the system couldn't reach the required temperature in the chamber within 10 minutes. To resolve this, simply perform a new cycle.
  • E
    Evelyn SmithSep 22, 2025
    What to do for 'Supply water error' on Tuttnauer 2840 ELV-D Laboratory Equipment?
    • S
      Stephen MooneySep 22, 2025
      If your Tuttnauer Laboratory Equipment displays 'Supply water error (digitat input option)', it indicates a malfunction in the city (tap) water supply. To resolve this, check and fix the city (tap) water supply.
  • B
    Barbara SteinSep 9, 2025
    What does 'High Pressure' mean on my Tuttnauer Laboratory Equipment and how can I fix it?
    • L
      Lisa HardinSep 9, 2025
      The 'High Pressure' message on your Tuttnauer Laboratory Equipment indicates that the chamber pressure rose 4.2 psi-29 kPa above the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. Perform a new cycle.
  • J
    Jacob RodriguezSep 14, 2025
    What does 'Air Error' mean on my Tuttnauer Laboratory Equipment and how can I fix it?
    • R
      Rachel PetersenSep 14, 2025
      The 'Air Error' message on your Tuttnauer Laboratory Equipment indicates that the autoclave did not reach atmospheric pressure after 100 minutes at the end of the cycle. Wait until the autoclave reaches atmospheric pressure and perform a new cycle.
  • N
    nathan36Sep 3, 2025
    What does 'Low Pressure (Cooling)' mean on my Tuttnauer Laboratory Equipment and how can I resolve it?
    • F
      Fred McdonaldSep 3, 2025
      The 'Low Pressure (Cooling)' message on your Tuttnauer Laboratory Equipment means that the pressure in the chamber did not reach the preset pressure before the cooling stage began. Check and fix the compressed air supply.
  • S
    Shawn WalterSep 4, 2025
    What should I do if my Tuttnauer Laboratory Equipment shows a compressed air supply error?
    • O
      okleinSep 4, 2025
      If your Tuttnauer Laboratory Equipment displays a compressed air supply error, check and fix the compressed air supply.
  • J
    jeffersonryanSep 23, 2025
    What does 'Power Down' mean on my Tuttnauer Laboratory Equipment?
    • J
      James BennettSep 23, 2025
      The 'Power Down' message on your Tuttnauer Laboratory Equipment indicates that a power outage occurred during the cycle. Turn on the autoclave, wait until it is ready and reaches a safe condition, and then perform a new cycle.
  • C
    Christie McconnellAug 29, 2025
    What does 'Low Pressure (Cooling)' mean on Tuttnauer 2840 ELV-D?
    • C
      Craig RuizAug 29, 2025
      If your Tuttnauer Laboratory Equipment displays 'Low Pressure (Cooling)', it means the pressure in the chamber did not reach the preset pressure before initiating the cooling stage. To resolve this, check and fix the compressed air supply.

Tuttnauer 2840 ELV-D Specifications

General IconGeneral
BrandTuttnauer
Model2840 ELV-D
CategoryLaboratory Equipment
LanguageEnglish