1.1.1
Operation and Maintenance
Manual
Electronic Laboratory Autoclaves
Models 2840 ELV-D
ELV-D: standard autoclave
C: optional system for fast cooling
PV: optional vacuum pump
BH: optional bio hazard filtration
WR: water recycling system
F: optional fan for super-fast cooling
Tuttnauer Europe B.V.,
Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda,
The Netherlands Tel: 31 (0) 765423510, Fax: 31 (0) 765423540
Cat. No. MAN205-0468001EN Rev H

Table of Contents

Other manuals for Tuttnauer 2840 ELV-D

Questions and Answers

  • S
    scolemanAug 31, 2025
    What does 'No Water' mean on Tuttnauer 2840 ELV-D?
    • A
      andrew53Aug 31, 2025
      If your Tuttnauer Laboratory Equipment displays 'No Water', it means the electrode in the chamber did not sense water within the preset time. Here are the steps to troubleshoot: 1. Check and fix the mineral free water supply. 2. Check and clean the water inlet filter. 3. Clean the water level electrode.
  • G
    grace22Sep 8, 2025
    What does 'Low Pressure' mean on my Tuttnauer Laboratory Equipment and how to resolve it?
    • B
      Barbara LaneSep 9, 2025
      The 'Low Pressure' message on your Tuttnauer Laboratory Equipment indicates that the chamber pressure dropped below the sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. Perform a new cycle.
  • C
    Christie McconnellAug 29, 2025
    What does 'Low Pressure (Cooling)' mean on Tuttnauer 2840 ELV-D?
    • C
      Craig RuizAug 29, 2025
      If your Tuttnauer Laboratory Equipment displays 'Low Pressure (Cooling)', it means the pressure in the chamber did not reach the preset pressure before initiating the cooling stage. To resolve this, check and fix the compressed air supply.
  • A
    Amanda StevensSep 23, 2025
    What to do if the printer of my Tuttnauer Laboratory Equipment does not print?
    • S
      Sara FullerSep 23, 2025
      If the printer of your Tuttnauer Laboratory Equipment isn't printing, first ensure that the paper roll is inserted correctly. If the issue persists, try switching the machine off and then back on to restart it.
  • J
    Janet SmithSep 10, 2025
    What does 'High Pressure (Ending)' mean on my Tuttnauer Laboratory Equipment?
    • C
      courtneyhansenSep 10, 2025
      The 'High Pressure (Ending)' message on your Tuttnauer Laboratory Equipment indicates that the system could not reach atmospheric pressure ± 10kPa during the ending stage. Perform a new cycle.
  • J
    Jacob RodriguezSep 14, 2025
    What does 'Air Error' mean on my Tuttnauer Laboratory Equipment and how can I fix it?
    • R
      Rachel PetersenSep 14, 2025
      The 'Air Error' message on your Tuttnauer Laboratory Equipment indicates that the autoclave did not reach atmospheric pressure after 100 minutes at the end of the cycle. Wait until the autoclave reaches atmospheric pressure and perform a new cycle.
  • E
    Evelyn SmithSep 22, 2025
    What to do for 'Supply water error' on Tuttnauer 2840 ELV-D Laboratory Equipment?
    • S
      Stephen MooneySep 22, 2025
      If your Tuttnauer Laboratory Equipment displays 'Supply water error (digitat input option)', it indicates a malfunction in the city (tap) water supply. To resolve this, check and fix the city (tap) water supply.
  • M
    Michele HallAug 27, 2025
    What does 'Pressure Time Error' mean on Tuttnauer Laboratory Equipment?
    • N
      Nathan WintersAug 27, 2025
      If your Tuttnauer Laboratory Equipment displays 'Pressure Time Error', it means that the system cannot reach the required pressure conditions in the chamber, after preset time, during the air removal stage. To resolve this, verify that the autoclave is not overloaded.
  • J
    Julie PorterAug 26, 2025
    What does 'High Pressure' mean on Tuttnauer 2840 ELV-D?
    • G
      Gabrielle LarsonAug 26, 2025
      If your Tuttnauer Laboratory Equipment displays 'High Pressure', it means the chamber pressure rose 4.2 psi-29 kPa above sterilization pressure (134°C = 304 kPa, 121°C = 205 kPa) for 2 seconds during the sterilization stage. To resolve this, perform a new cycle.
  • A
    allenbethanyAug 25, 2025
    What does 'High Temp. (Ending)' mean on Tuttnauer 2840 ELV-D Laboratory Equipment?
    • C
      charlesclarkAug 26, 2025
      If your Tuttnauer Laboratory Equipment displays 'High Temp. (Ending)', it means the system couldn't reach the required temperature in the chamber within 10 minutes. To resolve this, simply perform a new cycle.

Tuttnauer 2840 ELV-D Specifications

General IconGeneral
BrandTuttnauer
Model2840 ELV-D
CategoryLaboratory Equipment
LanguageEnglish