Questions and Answers

  • S
    serranostephanieSep 13, 2025
    Why are wraps burned after using Tuttnauer 2340EA?
    • C
      Charles EricksonSep 14, 2025
      To address wraps coming out burned from your Tuttnauer Laboratory Equipment, check if the water sensor is dirty. Also, verify that the wall outlet voltage is not too high and ensure items are not touching the chamber walls or bottom. Load the autoclave according to the instructions.
  • M
    Meghan ReynoldsAug 25, 2025
    What to do when my Tuttnauer Laboratory Equipment displays ‘High Temp’?
    • D
      David HarrellAug 25, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Temp’, clean the air jet and check for proper voltage at the wall outlet.
  • M
    Michelle CollinsAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer 2340EA Laboratory Equipment?
    • R
      robertsmaryAug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.
  • R
    Robert RobertsAug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer Laboratory Equipment?
    • J
      Jason JohnsonAug 27, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • W
    William ParsonsSep 14, 2025
    Why are instruments rusting in my Tuttnauer 2340EA?
    • G
      Gina HarveySep 14, 2025
      If instruments are rusting after sterilization in your Tuttnauer Laboratory Equipment, first determine if it's true rust by using a pencil eraser to remove the brown spot. If there is no pitting, the instrument is just dirty. Use distilled water for sterilizing and ensure dissimilar metals are not in contact with each other.
  • P
    Penny SmithSep 23, 2025
    Why does Tuttnauer 2340EA Laboratory Equipment printer feed paper constantly?
    • A
      Ana NguyenSep 23, 2025
      If the printer on your Tuttnauer Laboratory Equipment gives paper feeds all the time when the machine is switched on, make sure the ‘feed button’ on the printer is not stuck.
  • D
    dianeshermanSep 7, 2025
    What to do if the keyboard of my Tuttnauer 2340EA does not respond?
    • E
      Eric AllenSep 7, 2025
      If the keyboard of your Tuttnauer Laboratory Equipment does not respond, press the STOP key for 2 seconds to abort any program that may still be running.
  • A
    Anna FrancisSep 8, 2025
    How to troubleshoot a Tuttnauer 2340EA Laboratory Equipment system that shows no response when turned on?
    • P
      Paula RosalesSep 8, 2025
      If your Tuttnauer Laboratory Equipment does not respond when you turn it on, try the following: First, check and fix all connections. If the issue persists, replace any faulty cables, connectors, or the power switch.
  • L
    Louis AdamsSep 23, 2025
    Why is my Tuttnauer 2340EA making a loud noise?
    • G
      glenn82Sep 23, 2025
      If you hear a loud noise coming from the reservoir of your Tuttnauer Laboratory Equipment, the sterilization temperature may be set too high. The maximum setting is 274ºF (134ºC).
  • M
    martinezmelissaAug 31, 2025
    How to clean a very dirty Tuttnauer 2340EA Laboratory Equipment chamber?
    • J
      josephwilsonAug 31, 2025
      To clean a very dirty chamber in your Tuttnauer Laboratory Equipment, use Chamber Brite ™. Very dirty machines may require two cleanings.

Summary

General Information

Control Panel and Interface

Error Messages and Safety Measures

Explains common error messages, warning messages, and associated safety measures.

Sterilization Programs

Program Overview and Usage

Details the four sterilization programs, their settings, and operational sequences.

Installation and Setup

Electrical Connection and Safety

Covers electrical connection requirements, safety, and compliance with regulations.

Chamber Pitch Adjustment

Details the crucial chamber pitch adjustment for older units and its impact on operation.

Preparation Before Sterilization

Instrument Cleaning and Preparation

Instructions on cleaning, rinsing, and drying instruments before sterilization.

Packaging and Loading Practices

Guidance on packaging methods, tray placement, and preventing issues like water accumulation.

Instrument Handling and Sterilization Tips

Advice on handling specific materials, preventing cross-contamination, and ensuring proper exposure.

Operating Instructions

Pre-Operation Checks and Setup

Steps before starting operation, including water filling, power connection, and program selection.

Cycle Execution and Monitoring

How to start a cycle, monitor its progress, and understand indicator lights.

Maintenance Instructions

Preventive and Scheduled Maintenance

Outlines daily, weekly, and periodic maintenance tasks for optimal performance.

Troubleshooting Guide

Machine Not Responding or Low Heat Issues

Solutions for unresponsive machines and 'Low Heat' error messages.

Water Level and Pressure Problems

Troubleshooting for 'Low Water' and 'Low Pres' messages, including door seal checks.

Temperature Errors and Printer Malfunctions

Addresses 'Low Temp'/'High Temp' errors and common printer operational issues.

Tuttnauer 2340EA Specifications

General IconGeneral
TypeAutoclave
Power1200 W
Frequency50/60 Hz
Chamber MaterialStainless Steel
Sterilization Time Range3 - 30 minutes
Warranty1 year
Cycle Time30 minutes
Sterilization Temperature Range121°C to 134°C