OPERATION
&
MAINTENANCE
MANUAL
Table-Top Autoclaves
models
1730, 2340, 2540, 3140, 3850, 3870 M & MK
1730MK Valueklave
Cat. No. MAN205-0007-000E Rev. K
Tuttnauer U.S.A., 25 Power Drive Hauppauge, NY, 11788, USA. Tel (631) 737 4850, (800) 624 5836, Fax: (631) 737 0720
Tuttnauer Europe b.v., Paardeweide 36, 4824 EH, Breda, P.O. Box 7191, 4800 GD Breda, Netherlands. +31/76-5423510,
Fax: +31/76-5423540

Other manuals for Tuttnauer 2340 MK

Questions and Answers

  • B
    bbartonAug 14, 2025
    How to troubleshoot a Tuttnauer 2340 MK when the POWER indicator light doesn't turn on?
    • B
      Brendan SandersAug 14, 2025
      If the POWER indicator light on your Tuttnauer Laboratory Equipment does not light up, several issues could be the cause. First, the main switch might be defective, in which case you should check and replace it if necessary. It could also be a problem with the "POWER" bulb itself, so try replacing it. Inspect the line from the entry point to the main switch for any damage and repair it. Another potential issue is an internal electrical problem within the unit, which requires further investigation. Additionally, the “Power” light may need to be replaced.
  • J
    Jonathan SavageAug 17, 2025
    What to do if steam escapes from the safety valve of my Tuttnauer 2340 MK Laboratory Equipment?
    • T
      Tara SmithAug 17, 2025
      If steam is escaping from the safety valve of your Tuttnauer Laboratory Equipment, activate the safety valve. If the leaking continues after activation, you will need to replace the safety valve.
  • J
    Jean McgeeAug 21, 2025
    Why won't water enter the chamber of my Tuttnauer Laboratory Equipment?
    • K
      Kayla TaylorAug 21, 2025
      If water is not entering the chamber of your Tuttnauer Laboratory Equipment, it could be due to a clog in the multi-purpose valve or the pipe. Follow instructions in para. 6.10 to resolve this issue.
  • C
    cathy53Aug 25, 2025
    What to do if the door handle of my Tuttnauer 2340 MK won't turn to open?
    • K
      Karen EdwardsAug 25, 2025
      If the door handle of your Tuttnauer Laboratory Equipment cannot be turned counter-clockwise to open, it may be because the door pin is set in a groove. Try turning the handle slightly in the closing direction (clockwise) and then attempt to open it again. If the problem persists, the door locking system may be stuck or the bellows damaged. Refer to “Pressure Door Lock System”. After opening the door, replace the bellows.
  • J
    jrogersAug 27, 2025
    Why is the 'HEAT' light off and there's no heating in my Tuttnauer Laboratory Equipment?
    • W
      William HuntAug 28, 2025
      If the “HEAT” light is not on and there is no heating in your Tuttnauer Laboratory Equipment, the faulty thermostat (B10) should be fixed or replaced.
  • A
    Amy BlackwellAug 30, 2025
    Why is the 'HEAT' light off and heating insufficient in my Tuttnauer 2340 MK?
    • K
      Kevin NelsonAug 31, 2025
      If the “HEAT” light is not on and the heating is insufficient in your Tuttnauer Laboratory Equipment, the faulty thermostat (B10) should be fixed or replaced.
  • N
    Nicholas TurnerSep 3, 2025
    Why is the 'HEAT' light off, but my Tuttnauer 2340 MK still operates?
    • T
      tylermikaylaSep 3, 2025
      If the “HEAT” light is not on but the autoclave operates in your Tuttnauer Laboratory Equipment, the faulty thermostat (B10) should be fixed or replaced.
  • S
    Shelley PrestonSep 7, 2025
    Why does the multi-purpose valve on my Tuttnauer 2340 MK turn backwards?
    • P
      Patricia EvansSep 7, 2025
      If the multi-purpose valve on your Tuttnauer Laboratory Equipment turns backwards, the internal spring in the multi-purpose valve has likely broken. Replace the multi-purpose valve.
  • S
    simmonsstanleySep 10, 2025
    What causes the timer on my Tuttnauer 2340 MK to not time down?
    • M
      Matthew DavisSep 10, 2025
      If the timer on your Tuttnauer Laboratory Equipment does not time down, the internal gearing has likely worn down. Replace the timer.

Summary

General Information and Warranty

Warranty Information

Details on the warranty coverage, limitations, and activation process for the autoclave.

Description of Components

Installation Instructions

Setup and Chamber Pitch Adjustment

Step-by-step guide for positioning, leveling, and preparing the autoclave for use.

Preparation Before Sterilization

Operation Procedures

Loading and Unloading the Device

Instructions for safely loading and unloading items into the autoclave chamber.

Service and Maintenance Instructions

Preventive and Scheduled Maintenance

Routine maintenance tasks to keep the autoclave in good working condition.

Cleaning Table Top Autoclaves

Troubleshooting Common Issues

Power Indicator Light Problems

Diagnosing and correcting issues with the power indicator light.

Heat Indicator Light Problems

Diagnosing and correcting issues with the heat indicator light.

Dry Indicator Light Problems

Diagnosing and correcting issues with the dry indicator light.

Water Entry Issues

Troubleshooting problems related to water not entering the chamber correctly.