Laboratory Vertical Steam Sterilizers
Including Fast Cooling System
models 3840, 3850, 3870
ELV/WR/C/PV/BH/F-D
Cat. No. MAN205-0448002EN Rev F
Tuttnauer Europe b.v., Hoeksteen 11, 4815 PR, Breda, P.O. Box 7191, 4800 GD
Breda, Netherlands. +31/76-5423510, Fax: +31/76-5423540

Table of Contents

Questions and Answers

  • F
    Frank WhiteAug 19, 2025
    What to do if 'Low Vacuum' is displayed on Tuttnauer Laboratory Equipment?
    • L
      lorievansAug 19, 2025
      If the Tuttnauer Laboratory Equipment displays 'Low Vacuum', this indicates that a vacuum level of 15kPa was not reached within 20 minutes after the cycle started during the air removal stage. To resolve this, try performing a new cycle. If the issue persists, it may be necessary to call a technician, or the bio-hazard filter may be clogged.
  • F
    Frank ReynoldsAug 22, 2025
    What does 'Supplied Distilled Water Error' mean on Tuttnauer 3840 ELV Laboratory Equipment?
    • M
      melendezstephanieAug 22, 2025
      If the Tuttnauer Laboratory Equipment displays the message "Supplied Distilled Water Error", it means the mineral-free water reservoir is empty. To fix this, simply fill the mineral-free water reservoir.
  • R
    Richard BlackwellAug 25, 2025
    What to do if Tuttnauer Laboratory Equipment displays 'No Water'?
    • H
      Hayden HartAug 25, 2025
      If the Tuttnauer Laboratory Equipment displays 'No Water', first check and fix the mineral-free water supply. Then, inspect and clean the water inlet filter, and finally, clean the water level electrode.
  • M
    Michelle GrimesAug 27, 2025
    What to do when Tuttnauer 3840 ELV displays 'Cycle counter exceeded -'?
    • J
      Juan WagnerAug 27, 2025
      If the Tuttnauer Laboratory Equipment displays 'Cycle counter exceeded -', call for service. This is because the number of cycles since the last periodical maintenance has exceeded the preset number.
  • K
    Katherine ValdezAug 28, 2025
    What to do if Tuttnauer 3840 ELV Laboratory Equipment displays 'Cycle Failed'?
    • J
      Joshua DavisAug 28, 2025
      If the Tuttnauer Laboratory Equipment displays 'Cycle Failed', perform a new cycle.
  • M
    Mrs. Colleen SandovalAug 30, 2025
    How to troubleshoot a Tuttnauer Laboratory Equipment when the display is not activated?
    • A
      Anthony LunaAug 30, 2025
      If the display of your Tuttnauer Laboratory Equipment is not activated, first, check the power source to ensure it meets the required specifications. Next, verify that the ON/OFF switch is in the ‘On’ position. Also, make sure the power cord is properly connected to both the machine and the mains. Check the reset button on the cut-out thermostat and reset it if necessary. Finally, check if the circuit breaker has tripped and reset it if needed.
  • W
    Wayne MendozaSep 1, 2025
    What to do if my Tuttnauer Laboratory Equipment is leaking at the door?
    • B
      brianmillerSep 1, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, ensure that the door is tightened sufficiently. If the leakage persists, consider replacing the door gasket.
  • T
    Theresa BurgessSep 3, 2025
    How to troubleshoot a Tuttnauer 3840 ELV Laboratory Equipment machine that is not responding?
    • C
      Courtney ThompsonSep 4, 2025
      If the Tuttnauer Laboratory Equipment machine isn't responding, first ensure the main switch is in the ‘On’ position. Then, check that the power cord is properly connected to both the machine and the mains. Finally, lift the circuit breaker lever if it has tripped. These issues may be caused by the main switch being off, a disconnected power cord, or a tripped circuit breaker.
  • J
    Johnny AyersSep 5, 2025
    What to do if my Tuttnauer 3840 ELV Laboratory Equipment printer does not print?
    • J
      Jennifer McgeeSep 6, 2025
      If the Tuttnauer Laboratory Equipment printer isn't printing, make sure the paper roll is correctly inserted. If there's still no printing, switch the machine off and then back on to restart it. This could be happening because no paper is inserted or for no obvious reason.
  • J
    Jared CarrollSep 7, 2025
    Why is my Tuttnauer 3840 ELV Laboratory Equipment leaking at the door?
    • R
      Robin LesterSep 8, 2025
      If the Tuttnauer Laboratory Equipment is leaking at the door, the door gasket may be dirty or damaged. Try cleaning the door gasket. If the leaking continues, call for service.

Tuttnauer 3840 ELV Specifications

General IconGeneral
Model3840 ELV
CategoryLaboratory Equipment
Number of Trays2
DoorAutomatic
MaterialStainless Steel
Power Supply230V, 50/60Hz
Power1.5 kW
Sterilization Temperature Range105°C to 137°C
StandardsASME
Safety Featuresdoor lock
Chamber Volume85 L (22.5 Gallons)