What to do if the Tuttnauer 3840 Laboratory Equipment shows a 'No Water' error?
- BBarbara ProctorSep 6, 2025
If the Tuttnauer Laboratory Equipment displays 'No Water' message, it means that the electrode in the chamber did not sense water within the preset time. To solve this: 1. Check and fix the mineral-free water supply. 2. Check and clean the water inlet filter. 3. Clean the water level electrode.