15.4 Cleaning water strainer
Caution!
Before proceeding, Make sure that the electric cord is disconnected
and there is no pressure in the autoclave.
1. Open the strainer cover (see the rear view).
2. Remove the strainer element.
3. Rinse the strainer with water. Use a brush if necessary.
4. Reinstall the strainer element.
5. Close the strainer cover.
Caution!
Do not touch the strainer’s cover, mounted on the exhaust line, during
and short after operation.
Touching the hot strainer’s cover may cause severe injuries.
Cap
Gasket
Strainer
element
Strainer
Housing