Page 34 of 110 Pages
6.2 Program 2: Plastic
For plastic and other delicate instruments, when the manufacturer
recommends autoclaving at temperatures of 121C (250F).
Nominal parameters default settings
Sterilization temperature: 121C (250F)
Sterilization time: 15 minutes
Operations sequence:
(PV, pre-vacuum model only) Air removal stage; four vacuum pulses
removes air by drawing the chamber down to 20kPa.
Water enters the chamber and is heated until pressure reaches 110 kPa
Heating by actuation of electrical heaters until the sterilization
temperature is reached.
Sterilization temperature is maintained constant during the sterilization
time.
Fast exhaust, the steam is exhausted out of the chamber at a fast rate
until pressure decreases to ambient pressure.
Note:
The sterility of instruments processed in unwrapped cycles cannot be
maintained if exposed to non-sterile environment.