Operation and
Maintenance Manual
Electronic Laboratory Autoclaves
Models 3840, 3850, 3870 ELVG-D
ELVG-D: standard autoclave
C: optional system for fast cooling
PV: optional vacuum pump
BH: optional bio hazard filtration
F: optional fan for super-fast cooling
Cat. No. MAN205-0448004EN Rev H
Tuttnauer Europe B.V.,
Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda,
The Netherlands Tel: 31 (0) 765423510, Fax: 31 (0) 765423540

Table of Contents

Other manuals for Tuttnauer 3870 ELVG-D

Questions and Answers

  • D
    Denise GutierrezSep 9, 2025
    How to fix Tuttnauer 3870 ELVG-D printer that doesn't print anything on the paper?
    • V
      Vincent SolisSep 9, 2025
      If your Tuttnauer Laboratory Equipment printer prints, but nothing appears on the paper, it likely means the paper roll isn't installed correctly. Ensure you install the paper roll the right way, noting that only one side of the paper is printable.
  • J
    Jessica HarrisAug 25, 2025
    What to do if my Tuttnauer 3870 ELVG-D Laboratory Equipment displays 'Heat Time Error'?
    • K
      Kathleen SmithAug 25, 2025
      If your Tuttnauer Laboratory Equipment displays 'Heat Time Error', it means the system couldn't reach the required temperature in the chamber within the preset time. Try performing a new cycle. If the problem continues, call a technician.
  • L
    Lauren RiveraAug 25, 2025
    What does 'Low Temp' mean on my Tuttnauer Laboratory Equipment?
    • T
      Tanya SchultzAug 26, 2025
      If your Tuttnauer Laboratory Equipment displays 'Low Temp', it indicates that the temperature dropped below the sterilization temperature for more than 1 second during the sterilization cycle. Try performing a new cycle.
  • R
    Rachael RobertsSep 23, 2025
    Why is the air detector failing in my Tuttnauer Laboratory Equipment?
    • D
      Diana ZhangSep 23, 2025
      If the air detector fails, consider the following possible causes: * The steam quality might not be proper. Check the steam quality and non-condensable gases. * The induced air valve might be open or leaking. Close or repair the induced air valve.
  • J
    jamesmannSep 12, 2025
    What to do if my Tuttnauer 3870 ELVG-D Laboratory Equipment printer does not print?
    • R
      Robert HamptonSep 12, 2025
      If the printer on your Tuttnauer Laboratory Equipment isn't printing, first ensure that the paper roll is properly inserted. If it still doesn't print and there's no obvious reason, try switching the machine off and then back on to restart it.
  • A
    Amy NguyenSep 16, 2025
    Why are my Tuttnauer Laboratory Equipment sterilization results unsatisfactory?
    • S
      Samantha PerrySep 17, 2025
      If you're experiencing unsatisfactory sterilization results, consider the following possible causes: * There might be incomplete removal of air pockets. * Sterilization conditions might involve improper temperature or pressure being too low, or the sterilization time being too short. Set the correct temperature and time according to the type and quantity of material. * Incorrect loading of the autoclave may be trapping air in the load.
  • C
    Cheryl DominguezSep 4, 2025
    What does 'RTC Error - Please Set Current Date and Time' mean on my Tuttnauer Laboratory Equipment?
    • N
      Nichole HardinSep 4, 2025
      If your Tuttnauer Laboratory Equipment displays 'RTC Error - Please Set Current Date and Time', it means you need to set the current date and time. If the problem persists after setting the date and time, call a technician.
  • L
    Laura WilliamsSep 5, 2025
    What does 'Compresse d air supply error (ELVCPVG- D only)' mean on Tuttnauer 3870 ELVG-D?
    • M
      Mr. Bryan JonesSep 5, 2025
      If your Tuttnauer Laboratory Equipment displays 'Compresse d air supply error (ELVCPVG- D only)', it indicates a compressed air supply malfunction. Check and fix the compressed air supply.
  • X
    xjohnsonSep 18, 2025
    What to do if Tuttnauer Laboratory Equipment control unit does not initialize?
    • B
      brittanyphillipsSep 18, 2025
      If the control unit doesn't initialize, showing no display or lights on the front panel, it could be due to: * The control circuit breaker being switched off. Check and switch on the circuit breaker. * Defective power switches on the front panels of the autoclave or steam generator. Replace the defective switch.
  • A
    Angel DanielAug 30, 2025
    What does 'High Pressure (Ending)' mean on my Tuttnauer 3870 ELVG-D?
    • B
      Brett BowenAug 30, 2025
      If your Tuttnauer Laboratory Equipment displays 'High Pressure (Ending)', it means the system couldn't reach atmospheric pressure ± 10kPa during the ending stage. Perform a new cycle.

Summary

Safety Instructions

General Safety Guidelines

General safety guidelines and warnings for operating the autoclave.

Technical Data

Electrical Data

Electrical specifications for 230V and 400V configurations of the autoclave.

Utilities

Information on electrical power, compressed air, water, and drain requirements.

Safety Features

Overview of built-in safety features such as error display and relief valves.

Control Panel

Displayed Error Messages

Categorization of failures and reference to troubleshooting for error messages.

Sterilization Programs

Program 1: Glass

Details on the Glass sterilization program, including parameters and sequence.

Program 2: Plastic

Details on the Plastic sterilization program, including parameters and sequence.

Program 9: Bio Hazard 1

Details on the Bio Hazard 1 program for bio-hazard models.

Program 12: Vacuum Test

Details on the Vacuum Test program for pre-vacuum models.

Screens

Aborted Cycle Screens

Screens indicating aborted cycles after successful sterilization phase.

Fail Cycle Screens

General description of screens displayed during a failed cycle.

Checking and Changing Parameters

Changing a Parameter

Step-by-step guide on how to modify a specific parameter value.

System Parameters

Listing and explanation of system-wide parameters affecting all cycles.

Preparation before Sterilization

Operating Instructions

Service and Maintenance Instructions

Preventive and Scheduled Maintenance

Overview of maintenance operations to keep the autoclave functional.

Troubleshooting

Troubleshooting Chart

Chart to help users solve minor malfunctions before contacting service.

Tuttnauer 3870 ELVG-D Specifications