OPERATION
&
MAINTENANCE
MANUAL
Table-Top Autoclaves
models
1730, 2340, 2540, 3140, 3850, 3870 M & MK
1730MK Valueklave
Cat. No. MAN205-0007-000E Rev. K
Tuttnauer U.S.A., 25 Power Drive Hauppauge, NY, 11788, USA. Tel (631) 737 4850, (800) 624 5836, Fax: (631) 737 0720
Tuttnauer Europe b.v., Paardeweide 36, 4824 EH, Breda, P.O. Box 7191, 4800 GD Breda, Netherlands. +31/76-5423510,
Fax: +31/76-5423540

Other manuals for Tuttnauer 3850 M

Questions and Answers

Summary

General Information and Warranty

Product Warranty Details

Details the warranty coverage for defects in materials and workmanship for one year.

Component Identification and Functionality

Installation Procedures and Setup

Electrical Connection and Safety

Ensures proper electrical connection, compliance with safety rules and use of earth leakage relay.

Autoclave Setup and Chamber Pitch Adjustment

Steps for setting up the autoclave, including leveling and adjusting chamber pitch with water.

Water Quality Requirements for Sterilization

Acceptable Contaminant Levels in Water

Table listing physical characteristics and maximum acceptable contaminant levels for sterilizer water.

Preparation Before Sterilization

Packaging, Wrapping, and Loading Guidelines

Covers packaging materials, instrument cleaning, loading, and preparation best practices.

Autoclave Operation Guide

Loading, Unloading, and Reservoir Filling

Procedures for loading/unloading instruments and filling the water reservoir with distilled water.

Cycle Operation and Temperature Setting

Covers starting the cycle, setting temperature, pressure gauge, and utilizing the sterilization time table.

Service and Maintenance Procedures

Preventive and Scheduled Maintenance Overview

Describes daily, weekly, and periodic maintenance tasks to keep the autoclave in good working condition.

Cleaning the Air Jet

Procedure for cleaning the air jet, crucial for proper sterilization and preventing failed tests.

Checking the Safety Valve Function

Monthly procedure to check the safety valve for blockages and proper steam release.

Cleaning with Chamber Brite

Chamber Brite Cleaning Procedure

Step-by-step guide for cleaning the autoclave using Chamber Brite, including cautions.

Troubleshooting Common Issues

Unit Not Heating or Drying Properly

Addressing problems where the unit does not heat up or the dry cycle is ineffective.

Failed Spore Tests or Indicator Strips

Identifying reasons for failed tests, such as clogged air jets or incorrect loading.

Poor Drying of Wrapped Items

Addressing issues with wrapped or bagged items not drying properly after the cycle.

Failure to Reach Operating Pressure

Checking voltage, water levels, door seal, and safety valve for pressure issues.

Items Burning or Melting in Chamber

Investigating causes like incorrect water fill, items touching walls, or excessive time.

Tuttnauer 3850 M Specifications

General IconGeneral
BrandTuttnauer
Model3850 M
CategoryLaboratory Equipment
LanguageEnglish