EasyManua.ls Logo

Tuttnauer 3850E - User Manual

Tuttnauer 3850E
57 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
OPERATION
&
MAINTENANCE
MANUAL
Electronic Tabletop Autoclaves
Models 1730, 2340, 2540, 3140, 3850, 3870
E, EK, EA & EKA

Other manuals for Tuttnauer 3850E

Questions and Answers

  • E
    Ernest MackSep 14, 2025
    Why are wraps burned after using Tuttnauer 3850E?
    • R
      Richard NorrisSep 14, 2025
      To address wraps coming out burned from your Tuttnauer Laboratory Equipment, check if the water sensor is dirty. Also, verify that the wall outlet voltage is not too high and ensure items are not touching the chamber walls or bottom. Load the autoclave according to the instructions.
  • S
    Stephanie HarrisonAug 25, 2025
    What to do when my Tuttnauer 3850E Laboratory Equipment displays ‘High Temp’?
    • M
      mendezjosephAug 25, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Temp’, clean the air jet and check for proper voltage at the wall outlet.
  • H
    Holly VelasquezAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer 3850E?
    • M
      Melanie GutierrezAug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.
  • J
    Jessica WilliamsAug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer 3850E?
    • J
      Julia CarpenterAug 26, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • K
    Kelsey MillerSep 14, 2025
    Why are instruments rusting in my Tuttnauer 3850E Laboratory Equipment?
    • E
      Elijah JonesSep 15, 2025
      If instruments are rusting after sterilization in your Tuttnauer Laboratory Equipment, first determine if it's true rust by using a pencil eraser to remove the brown spot. If there is no pitting, the instrument is just dirty. Use distilled water for sterilizing and ensure dissimilar metals are not in contact with each other.
  • P
    patelheatherSep 23, 2025
    Why does Tuttnauer Laboratory Equipment printer feed paper constantly?
    • M
      markjordanSep 23, 2025
      If the printer on your Tuttnauer Laboratory Equipment gives paper feeds all the time when the machine is switched on, make sure the ‘feed button’ on the printer is not stuck.
  • D
    Deanna MoranSep 7, 2025
    What to do if the keyboard of my Tuttnauer 3850E Laboratory Equipment does not respond?
    • S
      Steven MayoSep 7, 2025
      If the keyboard of your Tuttnauer Laboratory Equipment does not respond, press the STOP key for 2 seconds to abort any program that may still be running.
  • S
    smithwilliamSep 8, 2025
    How to troubleshoot a Tuttnauer 3850E system that shows no response when turned on?
    • R
      russell51Sep 8, 2025
      If your Tuttnauer Laboratory Equipment does not respond when you turn it on, try the following: First, check and fix all connections. If the issue persists, replace any faulty cables, connectors, or the power switch.
  • A
    Allison CobbSep 23, 2025
    Why is my Tuttnauer 3850E Laboratory Equipment making a loud noise?
    • J
      Johnny GarciaSep 23, 2025
      If you hear a loud noise coming from the reservoir of your Tuttnauer Laboratory Equipment, the sterilization temperature may be set too high. The maximum setting is 274ºF (134ºC).
  • M
    Mary HurleyAug 31, 2025
    How to clean a very dirty Tuttnauer 3850E chamber?
    • B
      Bryce SullivanAug 31, 2025
      To clean a very dirty chamber in your Tuttnauer Laboratory Equipment, use Chamber Brite ™. Very dirty machines may require two cleanings.

Summary

GENERAL

Incoming Inspection

Inspect the autoclave for damage upon arrival and document any findings.

Warranty and Statement

Details warranty coverage, activation process, and repair procedures.

TECHNICAL DATA

Electrical Data

Provides electrical specifications and fuse/breaker ratings.

Specifications

Details overall dimensions and physical specifications.

KEYBOARD AND DISPLAY

Control Panel Keyboard Functions

Details the purpose and operation of each key on the control panel.

Displayed Messages and Safety Measures

Explains various messages and associated safety measures displayed by the unit.

STERILIZATION PROGRAMS

Program Details

Covers programs for unwrapped, wrapped items, glassware, and dry only.

PREPARATION BEFORE STERILIZATION

OPERATING INSTRUCTIONS

Initial Operation and Program Selection

Guides for first-time use, program selection, and starting a cycle.

MAINTENANCE INSTRUCTIONS

Scheduled Maintenance Tasks

Outlines daily, weekly, and periodic maintenance procedures.

TROUBLESHOOTING GUIDE

Machine Errors and Unresponsiveness

Solutions for low heat, low water, low pressure, low temperature, and unresponsiveness.

Tuttnauer 3850E Specifications

General IconGeneral
BrandTuttnauer
Model3850E
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals