OPERATION
&
MAINTENANCE
MANUAL
Electronic Tabletop Autoclaves
Models 1730, 2340, 2540, 3140, 3850, 3870
E, EK, EA & EKA

Other manuals for Tuttnauer 3850E

Questions and Answers

  • A
    alunaSep 2, 2025
    Why is my Tuttnauer 3850E leaking at the door?
    • A
      allisonwhiteSep 2, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, ensure the door is tightened enough and the door gasket is sealing the chamber. If necessary, replace the door gasket.
  • E
    Ernest MackSep 14, 2025
    Why are wraps burned after using Tuttnauer 3850E?
    • R
      Richard NorrisSep 14, 2025
      To address wraps coming out burned from your Tuttnauer Laboratory Equipment, check if the water sensor is dirty. Also, verify that the wall outlet voltage is not too high and ensure items are not touching the chamber walls or bottom. Load the autoclave according to the instructions.
  • M
    Mary HurleyAug 31, 2025
    How to clean a very dirty Tuttnauer 3850E chamber?
    • B
      Bryce SullivanAug 31, 2025
      To clean a very dirty chamber in your Tuttnauer Laboratory Equipment, use Chamber Brite ™. Very dirty machines may require two cleanings.
  • B
    Brittney ThomasSep 16, 2025
    Why does Tuttnauer 3850E take so long to heat up?
    • P
      Patrick NortonSep 16, 2025
      If your Tuttnauer Laboratory Equipment takes too long to heat up, clean the air jet and check for proper voltage at the wall outlet.
  • M
    Michael BlackwellSep 15, 2025
    What to do if spore tests fail in Tuttnauer 3850E?
    • J
      Jerome LewisSep 15, 2025
      If spore tests are failing or indicator strips are not turning in your Tuttnauer Laboratory Equipment, clean the air jet. Also, make sure the sterilization time and temperature are set correctly, and ensure the autoclave is not too heavily loaded.
  • P
    patelheatherSep 23, 2025
    Why does Tuttnauer Laboratory Equipment printer feed paper constantly?
    • M
      markjordanSep 23, 2025
      If the printer on your Tuttnauer Laboratory Equipment gives paper feeds all the time when the machine is switched on, make sure the ‘feed button’ on the printer is not stuck.
  • T
    Tara SmithAug 27, 2025
    How to troubleshoot Tuttnauer 3850E machine not responding?
    • E
      edavisAug 28, 2025
      If your Tuttnauer Laboratory Equipment machine isn't responding, verify that the main switch is in the ‘On’ position, the power cord is properly connected to both the machine and the mains, and the circuit breaker hasn't tripped. Also, check the reset button on the cut-out thermostat.
  • B
    Brenda SalazarAug 27, 2025
    What to do if Tuttnauer 3850E process fails during heating stage with a HIGH TEMP message?
    • B
      Brent RomeroAug 27, 2025
      If your Tuttnauer Laboratory Equipment process fails while initiating the heating stage and displays a HIGH TEMP message, try changing the setting of the thermostat.
  • J
    Jessica WilliamsAug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer 3850E?
    • J
      Julia CarpenterAug 26, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • H
    Holly VelasquezAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer 3850E?
    • M
      Melanie GutierrezAug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.

Summary

GENERAL

Incoming Inspection

Inspect the autoclave for damage upon arrival and document any findings.

Warranty and Statement

Details warranty coverage, activation process, and repair procedures.

TECHNICAL DATA

Electrical Data

Provides electrical specifications and fuse/breaker ratings.

Specifications

Details overall dimensions and physical specifications.

KEYBOARD AND DISPLAY

Control Panel Keyboard Functions

Details the purpose and operation of each key on the control panel.

Displayed Messages and Safety Measures

Explains various messages and associated safety measures displayed by the unit.

STERILIZATION PROGRAMS

Program Details

Covers programs for unwrapped, wrapped items, glassware, and dry only.

PREPARATION BEFORE STERILIZATION

OPERATING INSTRUCTIONS

Initial Operation and Program Selection

Guides for first-time use, program selection, and starting a cycle.

MAINTENANCE INSTRUCTIONS

Scheduled Maintenance Tasks

Outlines daily, weekly, and periodic maintenance procedures.

TROUBLESHOOTING GUIDE

Machine Errors and Unresponsiveness

Solutions for low heat, low water, low pressure, low temperature, and unresponsiveness.

Tuttnauer 3850E Specifications

General IconGeneral
BrandTuttnauer
Model3850E
CategoryLaboratory Equipment
LanguageEnglish