OPERATION
&
MAINTENANCE
MANUAL
Electronic Tabletop Autoclaves
Models 1730, 2340, 2540, 3140, 3850, 3870
E, EK, EA & EKA

Other manuals for Tuttnauer 3850E

Questions and Answers

  • H
    Holly VelasquezAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer 3850E?
    • M
      Melanie GutierrezAug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.
  • J
    Jessica WilliamsAug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer 3850E?
    • J
      Julia CarpenterAug 26, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • B
    Brenda SalazarAug 27, 2025
    What to do if Tuttnauer 3850E process fails during heating stage with a HIGH TEMP message?
    • B
      Brent RomeroAug 27, 2025
      If your Tuttnauer Laboratory Equipment process fails while initiating the heating stage and displays a HIGH TEMP message, try changing the setting of the thermostat.
  • T
    Tara SmithAug 27, 2025
    How to troubleshoot Tuttnauer 3850E machine not responding?
    • E
      edavisAug 28, 2025
      If your Tuttnauer Laboratory Equipment machine isn't responding, verify that the main switch is in the ‘On’ position, the power cord is properly connected to both the machine and the mains, and the circuit breaker hasn't tripped. Also, check the reset button on the cut-out thermostat.
  • M
    Michael StevensAug 28, 2025
    What should I do if my Tuttnauer 3850E Laboratory Equipment shows ‘Low Pres’?
    • C
      Carla GrahamAug 28, 2025
      If the 'Low Pres' message is displayed on your Tuttnauer Laboratory Equipment, clean the water level electrode inside the vessel. Also, ensure the door is fully closed, the door gasket is properly seated, and there is no steam leakage. Check the leveling of the machine.
  • L
    Lisa WilliamsAug 29, 2025
    How to fix ‘High Pres’ error on Tuttnauer 3850E Laboratory Equipment?
    • S
      seth41Aug 29, 2025
      If the 'High Pres' message appears on your Tuttnauer Laboratory Equipment, the air jet might be clogged; clean it. It could also mean that the heating elements are continuously on, or the temperature sensor is faulty. In these cases, contact a technician.
  • M
    melissa34Aug 30, 2025
    How to unclog the drain on Tuttnauer 3850E?
    • J
      Joshua RobertsAug 30, 2025
      If the drain of your Tuttnauer Laboratory Equipment is clogged: 1. Open the drain by turning it counterclockwise and place a heavy object over the reservoir cover. 2. Blow compressed air into the drain to force debris back into the reservoir, then clean out the reservoir. 3. Alternatively, disassemble the drain valve, clean it, and blow out the line. Reassemble the valve, ensuring the large 'O' ring is properly seated.
  • E
    Elizabeth RoseAug 23, 2025
    What should I do if my Tuttnauer 3850E displays ‘High Pres’?
    • K
      Kristen JacksonAug 23, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the air jet. Ensure the sterilization time and temperature are set correctly; if unsure, use the default settings. Also, make sure the autoclave is not too heavily loaded.
  • K
    Karen AllenSep 1, 2025
    What to do if my Tuttnauer 3850E Laboratory Equipment is not responding?
    • D
      deborah29Sep 1, 2025
      If your Tuttnauer Laboratory Equipment is not responding, ensure the main switch is in the ‘On’ position. Also, verify the power cord is properly connected to both the machine and the mains. Check the reset button on the cut-out thermostat and make sure the circuit breaker hasn't tripped; lift the circuit breaker lever if it has.
  • A
    alunaSep 2, 2025
    Why is my Tuttnauer 3850E leaking at the door?
    • A
      allisonwhiteSep 2, 2025
      If your Tuttnauer Laboratory Equipment is leaking at the door, ensure the door is tightened enough and the door gasket is sealing the chamber. If necessary, replace the door gasket.

Summary

GENERAL

Incoming Inspection

Inspect the autoclave for damage upon arrival and document any findings.

Warranty and Statement

Details warranty coverage, activation process, and repair procedures.

TECHNICAL DATA

Electrical Data

Provides electrical specifications and fuse/breaker ratings.

Specifications

Details overall dimensions and physical specifications.

KEYBOARD AND DISPLAY

Control Panel Keyboard Functions

Details the purpose and operation of each key on the control panel.

Displayed Messages and Safety Measures

Explains various messages and associated safety measures displayed by the unit.

STERILIZATION PROGRAMS

Program Details

Covers programs for unwrapped, wrapped items, glassware, and dry only.

PREPARATION BEFORE STERILIZATION

OPERATING INSTRUCTIONS

Initial Operation and Program Selection

Guides for first-time use, program selection, and starting a cycle.

MAINTENANCE INSTRUCTIONS

Scheduled Maintenance Tasks

Outlines daily, weekly, and periodic maintenance procedures.

TROUBLESHOOTING GUIDE

Machine Errors and Unresponsiveness

Solutions for low heat, low water, low pressure, low temperature, and unresponsiveness.

Tuttnauer 3850E Specifications

General IconGeneral
BrandTuttnauer
Model3850E
CategoryLaboratory Equipment
LanguageEnglish