Models 3840
3.7 Utilities
3.7.1 Electrical utility
Property
Value: 3850/3840
Value: 3870
Power supply
208/230VAC
±10%, 50Hz
400VAC ±10%,
50Hz
Recommended
circuit breaker
15A
10A
3.7.2 Other Utilities
Tap water
1/2", 2-3 Bar (29-44 psi)
Mineral free water
1/2", 2-3 Bar (29-44 psi)
Drain
2" Minimum Withstanding
temp. of 80°C (17F)
Attention:
A switch or circuit-breaker must be included in the building
installation. This switch or circuit-breaker shall be in close proximity
to the equipment, within easy reach of the operator; and marked as the
disconnecting device for the equipment.
The electrical net must be protected with a current leakage safety
relay.
The electrical network must comply with local rules or regulations.