OPERATION
&
MAINTENANCE
MANUAL
Electronic Table - Top Autoclaves
(With cooling option)
Models
2540, 3150, 3850, 3870, 5050, 5075 EL
Cat. No. MAN205-0035000EN Rev AA
Tuttnauer Europe B.V., Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda, The Netherlands
Tel: 31 (0) 765423510, Fax: 31 (0) 765423540
All manuals and user guides at all-guides.com

Table of Contents

Questions and Answers

  • D
    Darrell EvansAug 17, 2025
    What to do if Tuttnauer 3150 EL displays ‘Low Water’?
    • A
      Amanda HollowayAug 17, 2025
      If your Tuttnauer Laboratory Equipment displays a ‘Low Water’ message, you should: 1. Verify that there is water in the reservoir, and add water if needed. 2. Clean the water level electrode inside the vessel. 3. Check the leveling of the machine.
  • C
    curryjenniferAug 20, 2025
    What to do if Tuttnauer 3150 EL displays ‘Low Pres’?
    • S
      Shawn SmithAug 20, 2025
      If the Tuttnauer Laboratory Equipment displays ‘Low Pres’, verify that no instruments are in contact with the water level electrodes. Also, check the door for leakage and replace the door gasket if necessary.
  • C
    Charles LindseyAug 23, 2025
    What to do if Tuttnauer 3150 EL displays ‘High Pres’?
    • M
      Michael GeorgeAug 23, 2025
      If the Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the water level electrode inside the vessel. Also, check the levelling of the machine.
  • Z
    zbrownAug 26, 2025
    What to do if Tuttnauer Laboratory Equipment is not responding?
    • D
      donna54Aug 26, 2025
      If the Tuttnauer Laboratory Equipment is not responding, you should: 1. Check the outlet and verify that the voltage requirements have been met, including phase. 2. Ensure the main switch is in the ‘On’ position. 3. Confirm the power cord is properly connected to both the machine and the mains.
  • N
    Nicole McguireAug 28, 2025
    Why is Tuttnauer Laboratory Equipment leaking at the door?
    • T
      Troy SimmonsAug 28, 2025
      If the Tuttnauer Laboratory Equipment is leaking at the door, make sure the door is tightened enough. If the leaking continues, replace the door gasket.
  • J
    Jeremy FoleyAug 31, 2025
    What to do if Tuttnauer 3150 EL Laboratory Equipment displays ‘High Temp’?
    • R
      Rebecca CookAug 31, 2025
      If the Tuttnauer Laboratory Equipment displays ‘High Temp’, clean the water level electrode inside the vessel and check the leveling of the machine.
  • R
    robert60Sep 3, 2025
    What to do if Tuttnauer Laboratory Equipment displays ‘Low Heat’?
    • H
      Heidi GoodmanSep 3, 2025
      If the Tuttnauer Laboratory Equipment displays ‘Low Heat’, make sure the machine has the proper amount of sterilization load and is not overloaded.
  • C
    christopher74Sep 5, 2025
    How to fix Tuttnauer 3150 EL when the printer prints but nothing is on the paper?
    • R
      Ricardo HowellSep 5, 2025
      If the printer prints, but nothing appears on the paper, ensure the paper is mounted correctly, as only one side is printable. Also, make sure the paper is inserted in the printer. If the issue persists, switch off the machine and switch it back on while pressing the feed button on the printer. If a test printout appears, the printer is working, and there may be an electronics issue.
  • P
    Pam MercadoSep 8, 2025
    Why Tuttnauer 3150 EL Laboratory Equipment water does not exit chamber?
    • N
      Nicholas MartinezSep 8, 2025
      If water does not exit the chamber of your Tuttnauer Laboratory Equipment due to a clogged outlet strainer, clean the strainer according to the instructions.
  • J
    Jorge KnightSep 12, 2025
    What to do if Tuttnauer Laboratory Equipment printer does not print?
    • C
      Cody ThomasSep 12, 2025
      If the printer does not print, switch off the Tuttnauer Laboratory Equipment and switch it back on while pressing the feed button on the printer. If the printer still does not print the test printout, there is a problem with the printer.

Tuttnauer 3150 EL Specifications

General IconGeneral
TypeAutoclave
Power Supply230 V
Power Consumption3000 W
Control SystemMicroprocessor Controlled
Frequency50/60 Hz
Cycle Time30 min
Door TypeManual
MaterialStainless Steel
Safety FeaturesOver-temperature protection
Sterilization Temperature Range105°C to 138°C