Tuttnauer 3170 - User Manual

Tuttnauer 3170
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
OPERATION
&
MAINTENANCE
MANUAL
Laboratory Vertical Steam Sterilizers
models
2540, 3150, 3170, 3850, 3870, 5050, 5075
ELV
Standard Autoclave
ELVC
Including Fast Cooling System
ELVPRC
Including Preparation for Fast Cooling System
Cat. No. MAN205-0036000EN Rev AA
Tuttnauer Europe B.V., Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda, The Netherlands
Tel: 31 (0) 765423510, Fax: 31 (0) 765423540

Table of Contents

Other manuals for Tuttnauer 3170

Questions and Answers

  • M
    Mercedes CordovaAug 12, 2025
    What to do if ‘Low Pres’ is displayed on Tuttnauer 3170?
    • E
      Elizabeth ReevesAug 12, 2025
      If your Tuttnauer Laboratory Equipment displays ‘Low Pres’, check the door for any leakage and replace the door gasket if needed. Also, make sure that no instrument is touching the electrode by sticking out through the basket holes. Rearrange the instruments in the basket and restart the cycle.
  • M
    Michael Frazier PhDAug 16, 2025
    Why can't I change programs and the FAIL light remains on Tuttnauer 3170?
    • D
      Dr. Christy Flores MDAug 16, 2025
      If you cannot change programs and the FAIL light remains on for your Tuttnauer Laboratory Equipment, it might be due to a problem with the printer paper. Turn off the main switch. Remove the printer paper. Then, turn the main switch back on. The cycle should return to the previous stage. Insert the printer paper, making sure it is correctly positioned.
  • A
    adamsjoshuaAug 19, 2025
    How to troubleshoot a Tuttnauer Laboratory Equipment that is not responding?
    • J
      Johnny SmithAug 19, 2025
      If the Tuttnauer Laboratory Equipment is not responding, first, check the power source to ensure it meets the required specifications. Then, verify that the main switch is in the ‘On’ position. Ensure the power cord is securely connected to both the machine and the mains. Check the reset button on the cut-out thermostat and reset it if necessary. Finally, check the circuit breaker and reactivate it if it has tripped.
  • D
    Denise McdonaldAug 23, 2025
    Why the printer prints, but nothing is printed on the paper on Tuttnauer Laboratory Equipment?
    • K
      Katherine LeeAug 23, 2025
      If the Tuttnauer Laboratory Equipment printer prints but nothing appears on the paper, ensure the paper is mounted correctly, as only one side of the paper is printable.
  • R
    Rhonda AndrewsAug 27, 2025
    What to do if the printer does not print on Tuttnauer 3170?
    • W
      Wayne BassAug 27, 2025
      If the Tuttnauer Laboratory Equipment printer does not print, first ensure that the paper is inserted in the printer. Then, switch off the machine and switch it back on while pressing the feed button on the printer. If the printer prints a test printout, the printer is O.K. and there is a problem with the electronics. Contact your dealer to solve the problem. If the printer does not print the test printout, there is a problem with the printer. Contact your dealer to solve the problem.
  • R
    robertsanthonyAug 31, 2025
    Why does the Tuttnauer 3170 Laboratory Equipment feed paper all the time?
    • A
      Amanda CombsAug 31, 2025
      If the Tuttnauer Laboratory Equipment feeds paper all the time when switched on, make sure the ‘feed button’ on the printer is not stuck.
  • T
    timothy92Sep 3, 2025
    How to fix steam leaks at the door of Tuttnauer 3170 Laboratory Equipment?
    • S
      Stephanie MaddoxSep 3, 2025
      If steam leaks at the door of your Tuttnauer Laboratory Equipment, make sure the door is tightened enough and replace the door gasket.
  • S
    Sandra ColemanSep 7, 2025
    Why does the exhaust stage take a very long time on Tuttnauer 3170?
    • S
      sabrinaandradeSep 7, 2025
      If the exhaust stage takes a very long time when running a cycle on your Tuttnauer Laboratory Equipment, it is normal if you are running a ‘liquids’ program.

Summary

SAFETY INSTRUCTIONS

2.1 General Safety Precautions

Crucial safety guidelines for operating and maintaining the autoclave.

TECHNICAL DATA

3.12 Description of Operation

Step-by-step explanation of the heating, sterilization, and cooling cycles.

STERILIZATION PROGRAMS

KEYBOARD (keys and display)

5.1 Description and Functions of the Front Panel Keyboard

Explanation of the control panel keys and their functions.

5.3 Displayed Error Messages and Safety Measures

Identification and handling of error messages and associated safety measures.

PREPARATION BEFORE STERILIZATION

7.1 Instruments

Guidelines for preparing and loading instruments for sterilization.

7.3 Liquids

Guidelines for preparing and loading liquids for sterilization.

7.4 Loading

Information on loading baskets and their types.

OPERATION

8.1 Operating Instructions

Step-by-step guide for operating the autoclave.

8.3 Loading and Unloading the Device

Safety guidelines and procedures for loading and unloading the autoclave.

DOOR SAFETY SYSTEM

9.1 Solenoid locking device

Explanation of the solenoid mechanism that locks the door.

SERVICE AND MAINTENANCE

10.1 Preventive Maintenance

Daily, weekly, and periodic maintenance tasks for optimal performance.

TROUBLESHOOTING

11.1 General Troubleshooting Guide

A chart to help users resolve common malfunctions with the autoclave.

Tuttnauer 3170 Specifications

General IconGeneral
BrandTuttnauer
Model3170
CategoryLaboratory Equipment
LanguageEnglish