4.3 Program 3 (3 –Waste)
For waste materials when the manufacturer recommends autoclaving at a
temperature of 121ºC.
Nominal Parameters
Sterilization temp.: 121ºC (250ºF)
Sterilization time: 20 mins.
Operations Sequence
Heating phase; water enters the chamber and heats up until the
sterilization temperature is reached.
Sterilization phase; temperature is maintained constant at the
preset level for the sterilization time.
Slow exhaust (Ex. Mode =3); steam is slowly exhausted from the
chamber, until it reaches a temperature of 100ºC. At this point the
exhaust valve is opened.
The options available are described in paragraph 2. (Technical data).
TIME
PRESSURE (kPa)
TEMPERATURE