10.3 Replacing the Door Gasket
(For models 2540, 3150, 3170, 3850, 3870)
(Gaskets of models 5050, 5075 shall be replaced by a technician. See
technician manual)
Pull off the gasket from the door groove and install the new gasket
referring to the drawings as above points 1, 2 and 3.
Caution!
This gasket is designed with a trapezoidal cross section. The
gasket should be placed with the widest side towards the door.
10.4 Checking the Safety Valve
The safety valve is located on the rear side of the autoclave
In order to prevent the safety valve from a blockage, operate it once a
month.