10.1.4 Periodical Tests
1. Once every month activate the safety valve (see para. 10.4).
10.2 Replacing the Air Filter
In order to increase the pressure during the fast cooling stage, compressed
filtered air enters the chamber via a solenoid valve. The filtration of the air
is performed by the bacteriological filter that is placed at the inlet of the
chamber, through a solenoid valve.
The filter is mounted at the right side of the autoclave. .
To replace the filter proceed as follows:
1. Open the right service door of the autoclave.
2. Unscrew the fittings (upper and lower fitting) connecting the filter to
the piping.
3. Replace the filter with a new one and reconnect the fittings.
4. Close the service door.
Air filter