Page 54 of 58 Pages
11
TROUBLESHOOTING
This troubleshooting chart enables the user to solve minor malfunctions, prior to contacting our service department.
Only technical personnel having proper qualifications and holding technical documentation (including a technician
manual) and adequate information are authorized to service the apparatus.
Solution
1.1 Check the power source and verify that it is according to specifications.
1.2 Verify that the main switch is in the ‘On’ position.
(see top view drawing).
1.3 Make sure the power cord is properly connected to the machine and the mains.
(see rear view drawing)
1.4 Check the reset button on the cut-out thermostat. Reset if necessary.
1.5 Make sure the circuit breaker has not tripped and reactivate it if necessary.
2.1 Make sure the machine has the proper amount of sterilization load.
3.1 Check the door for leakage and replace the door gasket if necessary.
(see para. 10.3 Replacing the Door Gasket)
3.2 Verify that no instrument makes contact with the electrode by protruding
through the holes of the basket. Rearrange the instruments in the basket and
restart the cycle.
4.1 Clean the water level electrode inside the vessel.
5.1 Clean the water level electrode inside the vessel.
Problem
1. The machine is not responding
2. ‘Low Heat’ is displayed
3. ‘Low Pres’ is displayed
4. ‘High Temp’ is displayed
5. ‘High Pres’ is displayed