10.5 Cleaning water outlet strainer
Caution!
Before proceeding, Make sure that the electric cord is
disconnected and there is no pressure in the autoclave.
1. Open the left service door.
2. Open the strainer cover.
3. Remove the strainer element.
4. Rinse the strainer with water, using a brush if necessary.
5. Reinstall the strainer element.
6. Close the strainer cover.
7. Close the left service door.
Attention:
Ensure the gasket is not misplaced, as this will prevent using
the autoclave.
Strainer
Cap
Strainer
element
Strainer
housing
Silicon
gasket