Why is Tuttnauer Laboratory Equipment leaking at the door?
- RRoberto SnyderSep 9, 2025
If the Tuttnauer Laboratory Equipment is leaking at the door, make sure the door is tightened enough. If the leaking continues, replace the door gasket.
Why is Tuttnauer Laboratory Equipment leaking at the door?
If the Tuttnauer Laboratory Equipment is leaking at the door, make sure the door is tightened enough. If the leaking continues, replace the door gasket.
Why can't I change programs and the FAIL light remains on Tuttnauer Laboratory Equipment?
If you cannot change programs and the FAIL light remains on for your Tuttnauer Laboratory Equipment, it might be due to a problem with the printer paper. Turn off the main switch. Remove the printer paper. Then, turn the main switch back on. The cycle should return to the previous stage. Insert the printer paper, making sure it is correctly positioned.
What to do if Tuttnauer Laboratory Equipment displays ‘Low Water’?
If your Tuttnauer Laboratory Equipment displays a ‘Low Water’ message, you should: 1. Verify that there is water in the reservoir, and add water if needed. 2. Clean the water level electrode inside the vessel. 3. Check the leveling of the machine.
What to do if Tuttnauer 3150 Laboratory Equipment displays ‘Low Pres’?
If the Tuttnauer Laboratory Equipment displays ‘Low Pres’, verify that no instruments are in contact with the water level electrodes. Also, check the door for leakage and replace the door gasket if necessary.
What to do if Tuttnauer 3150 Laboratory Equipment displays ‘High Pres’?
If the Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the water level electrode inside the vessel. Also, check the levelling of the machine.
How to troubleshoot a Tuttnauer 3150 Laboratory Equipment that is not responding?
If the Tuttnauer Laboratory Equipment is not responding, first, check the power source to ensure it meets the required specifications. Then, verify that the main switch is in the ‘On’ position. Ensure the power cord is securely connected to both the machine and the mains. Check the reset button on the cut-out thermostat and reset it if necessary. Finally, check the circuit breaker and reactivate it if it has tripped.
What to do if Tuttnauer 3150 printer does not print?
If the printer does not print, switch off the Tuttnauer Laboratory Equipment and switch it back on while pressing the feed button on the printer. If the printer still does not print the test printout, there is a problem with the printer.
Why Tuttnauer Laboratory Equipment printer feeds paper continuously?
If the printer feeds paper continuously when the Tuttnauer Laboratory Equipment is switched on, make sure the ‘feed button’ on the printer is not stuck.
What to do if Tuttnauer 3150 is not responding?
If the Tuttnauer Laboratory Equipment is not responding, you should: 1. Check the outlet and verify that the voltage requirements have been met, including phase. 2. Ensure the main switch is in the ‘On’ position. 3. Confirm the power cord is properly connected to both the machine and the mains.
How to fix steam leaks at the door of Tuttnauer 3150?
If steam leaks at the door of your Tuttnauer Laboratory Equipment, make sure the door is tightened enough and replace the door gasket.
Details on receiving inspection for mechanical damage and manufacturer's warranty terms.
Critical safety guidelines and precautions for autoclave operation.
Device dimensions, operating conditions, electrical data, and standards.
Recommended for sterilizing unwrapped instruments at 134°C.
For instruments requiring autoclaving at 121°C with a drying cycle.
For sterilizing liquids at 121°C without a drying cycle.
For materials at 121°C with forced cooling to 85°C.
Explanation of front panel keys, display, and indicators for operation.
Steps for preparing instruments, packs, and liquids for sterilization.
Step-by-step guide for running the autoclave, loading, and unloading.
Daily, weekly, and periodic maintenance tasks to ensure optimal performance.
Procedures for draining reservoir, cleaning strainer, and checking safety valve.
Explanation of door locking mechanisms for different model series.
Solutions for common problems like machine not responding or error messages.
Specifications for stainless steel baskets and containers for waste products.
Lists of available spare parts and accessories with their catalog numbers.
| Number of Trays | 3 |
|---|---|
| Frequency | 50/60 Hz |
| Chamber Material | Stainless Steel |
| Tray Dimensions | 300 mm |
| Power Consumption | 1600W |