OPERATION
&
MAINTENANCE
MANUAL
Electronic Table - Top Autoclaves
(With cooling option)
Models
2540, 3150, 3850, 3870, 5050, 5075 EL
Cat. No. MAN205-0035000EN Rev AA
Tuttnauer Europe B.V., Hoeksteen 11 4815 PR P.O. Box 7191 4800 GD Breda, The Netherlands
Tel: 31 (0) 765423510, Fax: 31 (0) 765423540

Table of Contents

Other manuals for Tuttnauer 3150

Questions and Answers

  • T
    Taylor JensenSep 9, 2025
    Why is Tuttnauer Laboratory Equipment leaking at the door?
    • R
      Roberto SnyderSep 9, 2025
      If the Tuttnauer Laboratory Equipment is leaking at the door, make sure the door is tightened enough. If the leaking continues, replace the door gasket.
  • A
    adam11Aug 23, 2025
    Why can't I change programs and the FAIL light remains on Tuttnauer Laboratory Equipment?
    • B
      Brandi BaileyAug 23, 2025
      If you cannot change programs and the FAIL light remains on for your Tuttnauer Laboratory Equipment, it might be due to a problem with the printer paper. Turn off the main switch. Remove the printer paper. Then, turn the main switch back on. The cycle should return to the previous stage. Insert the printer paper, making sure it is correctly positioned.
  • L
    Lauren GonzalezAug 25, 2025
    What to do if Tuttnauer Laboratory Equipment displays ‘Low Water’?
    • T
      Teresa ParkAug 26, 2025
      If your Tuttnauer Laboratory Equipment displays a ‘Low Water’ message, you should: 1. Verify that there is water in the reservoir, and add water if needed. 2. Clean the water level electrode inside the vessel. 3. Check the leveling of the machine.
  • C
    Charles DoyleAug 26, 2025
    What to do if Tuttnauer 3150 Laboratory Equipment displays ‘Low Pres’?
    • J
      Jeffrey WilliamsAug 26, 2025
      If the Tuttnauer Laboratory Equipment displays ‘Low Pres’, verify that no instruments are in contact with the water level electrodes. Also, check the door for leakage and replace the door gasket if necessary.
  • M
    Michael DayAug 27, 2025
    What to do if Tuttnauer 3150 Laboratory Equipment displays ‘High Pres’?
    • M
      Margaret MillerAug 28, 2025
      If the Tuttnauer Laboratory Equipment displays ‘High Pres’, clean the water level electrode inside the vessel. Also, check the levelling of the machine.
  • B
    Barbara WalterAug 29, 2025
    How to troubleshoot a Tuttnauer 3150 Laboratory Equipment that is not responding?
    • J
      Jennifer LiuAug 29, 2025
      If the Tuttnauer Laboratory Equipment is not responding, first, check the power source to ensure it meets the required specifications. Then, verify that the main switch is in the ‘On’ position. Ensure the power cord is securely connected to both the machine and the mains. Check the reset button on the cut-out thermostat and reset it if necessary. Finally, check the circuit breaker and reactivate it if it has tripped.
  • N
    Nicholas ColeSep 18, 2025
    What to do if Tuttnauer 3150 printer does not print?
    • J
      josebakerSep 18, 2025
      If the printer does not print, switch off the Tuttnauer Laboratory Equipment and switch it back on while pressing the feed button on the printer. If the printer still does not print the test printout, there is a problem with the printer.
  • D
    David SalasSep 18, 2025
    Why Tuttnauer Laboratory Equipment printer feeds paper continuously?
    • M
      Michael EvansSep 22, 2025
      If the printer feeds paper continuously when the Tuttnauer Laboratory Equipment is switched on, make sure the ‘feed button’ on the printer is not stuck.
  • B
    Bianca HayesSep 7, 2025
    What to do if Tuttnauer 3150 is not responding?
    • M
      mathewsmichaelaSep 8, 2025
      If the Tuttnauer Laboratory Equipment is not responding, you should: 1. Check the outlet and verify that the voltage requirements have been met, including phase. 2. Ensure the main switch is in the ‘On’ position. 3. Confirm the power cord is properly connected to both the machine and the mains.
  • A
    alan56Sep 4, 2025
    How to fix steam leaks at the door of Tuttnauer 3150?
    • K
      knightashleySep 5, 2025
      If steam leaks at the door of your Tuttnauer Laboratory Equipment, make sure the door is tightened enough and replace the door gasket.

Summary

General Information and Safety

Incoming Inspection and Warranty

Details on receiving inspection for mechanical damage and manufacturer's warranty terms.

Safety Instructions

Critical safety guidelines and precautions for autoclave operation.

Technical Specifications

Device dimensions, operating conditions, electrical data, and standards.

Sterilization Programs

Program 1: Unwrapped Instruments

Recommended for sterilizing unwrapped instruments at 134°C.

Program 2: Instruments with Drying

For instruments requiring autoclaving at 121°C with a drying cycle.

Program 4: Liquids

For sterilizing liquids at 121°C without a drying cycle.

Program 6: Liquids with Cooling

For materials at 121°C with forced cooling to 85°C.

Operation and Control

Keyboard and Display Functions

Explanation of front panel keys, display, and indicators for operation.

Preparation Before Sterilization

Steps for preparing instruments, packs, and liquids for sterilization.

Operating Instructions

Step-by-step guide for running the autoclave, loading, and unloading.

Maintenance, Troubleshooting, and Parts

Preventive and Scheduled Maintenance

Daily, weekly, and periodic maintenance tasks to ensure optimal performance.

Routine Cleaning and Valve Checks

Procedures for draining reservoir, cleaning strainer, and checking safety valve.

Door Safety Systems

Explanation of door locking mechanisms for different model series.

Troubleshooting Common Issues

Solutions for common problems like machine not responding or error messages.

Baskets and Containers

Specifications for stainless steel baskets and containers for waste products.

Spare Parts and Accessories

Lists of available spare parts and accessories with their catalog numbers.

Tuttnauer 3150 Specifications

General IconGeneral
Number of Trays3
Frequency50/60 Hz
Chamber MaterialStainless Steel
Tray Dimensions300 mm
Power Consumption1600W

Related product manuals