released and valve opens, chamber is aerated and
residual pressure discharged from the chamber.
This way, the second step, opening the door by
pulling it, can be done safely, with no risk for the
operator, of steam or hot water burst.
After the ”door light” had signaled the operator will
press the “up” key and the solenoid will retract.
At this stage the unlocking operation is completed
and the operator can open the door.
9.8 Cleaning Table Top Autoclaves with Chamber Brite™
CHAMBER BRITE™ is a cleaning and
descaling agent designed specifically for
the cleaning and removal of water deposits,
oxides and other sediments that are found
in steam sterilizers. The material is a
combination of acidic salts and additional
cleaning materials.
Cleaning Procedure
1. Important – all steps in this procedure
must be completed without
Interruption.
2. When the autoclave chamber is cold,
remove instruments and tray from the
autoclave.
3. Open the door and spread the contents
of a packet in a straight even line
along the bottom of the chamber, from
back to front.
4. Select and start program 1. When the
cycle is finished exhaust the unit.
5. At the end of the exhaust cycle drain
the water from the reservoir.
6. Fill the water reservoir with distilled
water.
7 Repeat a sterilization cycle without
Chamber Brite™ powder, to remove
any excessive dirt in the pipes. Select
and start program 1. When the cycle is
finished exhaust the unit
8. At the end of the exhaust cycle drain
the water from the reservoir.
9. Turn the autoclave off and allow
chamber to cool.
10. Wipe the interior of the chamber with