Page 46 of 66 Pages
If the operator presses the START key and the door is not completely
closed, the process will not start and the
DOOR indicator will flash
twice then turn off and the buzzer will sound four times.
CAUTION
Do not touch the strainer’s cover, mounted on the exhaust line,
during and short after operation.
Touching the hot strainer’s cover may cause severe injuries.
NOTE
At first operation of models 5050, 5075, when operated from cold
start, a few milliliters of water may accumulate in the “C’ shape
frame of the door. This is normal and results from condensed
steam.
12. At the end of the cycle, the START light is put off, the END message
is displayed and the buzzer will ring a continuous sound for 7
seconds. In case of a failed cycle, the diagnosis of the failure
followed by the
FAIL!!!” message will be printed and the buzzer
will output an interrupted sound.
13. Open the door and unload the sterilized material from chamber.
13.1For models 2540, 3150, 3850, 3870, in case of fail press the
START/STOP key to cancel the locking of the door.
13.2For models 5050, 5075, to release the door locking at the end of
operation, press the
UP key. The same applies at power up after fail.
Rotate the handle counter-clockwise to pull out the locking arms from
the retaining brackets.
For unloading hints – see para. 7.2.3 (Unloading).
14. At the end of each working day close the main water valve.
8.2 Loading and unloading the Device
8.2.1 Safety
Protective equipment and clothes and other safety instructions
should be implemented in accordance with local and national
regulations and/or rules!
For proper sterilization - Do not overload the chamber. Only
autoclavable products shall be used; please refer to the
manufacturer instructions for sterilization of unknown materials
or instruments.
8.2.2 Loading
Correct loading of the autoclave is essential to successful
sterilizing for several reasons. Efficient air removal from the
chamber and the load will permit total steam penetration and
saturation, and allow proper drainage of condensate.
Additionally, correct loading will reduce damage to packs and
their contents and maximize efficient use of the sterilizer.
For detailed loading instructions, see para. 7 (Preparation before
sterilization)