  • W
    William MooreSep 14, 2025
    Why are wraps burned after using Tuttnauer 1730EK Laboratory Equipment?
    • A
      Andrea BeardSep 14, 2025
      To address wraps coming out burned from your Tuttnauer Laboratory Equipment, check if the water sensor is dirty. Also, verify that the wall outlet voltage is not too high and ensure items are not touching the chamber walls or bottom. Load the autoclave according to the instructions.
  • H
    harrisjohnnyAug 25, 2025
    What to do when my Tuttnauer 1730EK Laboratory Equipment displays ‘High Temp’?
    • B
      Brian TurnerAug 25, 2025
      If your Tuttnauer Laboratory Equipment displays ‘High Temp’, clean the air jet and check for proper voltage at the wall outlet.
  • R
    rbrowningAug 25, 2025
    What does ‘Low Heat’ mean on my Tuttnauer Laboratory Equipment?
    • B
      Brooke SullivanAug 26, 2025
      If the message ‘Low Heat’ is displayed during heat up on your Tuttnauer Laboratory Equipment, it indicates a bad temperature sensor. If heating elements are remaining on instead of cycling on and off, call for a technician.
  • T
    Tamara SmithAug 26, 2025
    How to clean a very dirty chamber in my Tuttnauer 1730EK?
    • D
      Daniel OwenAug 27, 2025
      If the chamber of your Tuttnauer Laboratory Equipment is very dirty, use Chamber Brite ™ to clean the autoclave; very dirty machines may require two cleanings.
  • D
    David MorrisonSep 14, 2025
    Why are instruments rusting in my Tuttnauer Laboratory Equipment?
    • S
      Shawn RodriguezSep 14, 2025
      If instruments are rusting after sterilization in your Tuttnauer Laboratory Equipment, first determine if it's true rust by using a pencil eraser to remove the brown spot. If there is no pitting, the instrument is just dirty. Use distilled water for sterilizing and ensure dissimilar metals are not in contact with each other.
  • X
    xsnowSep 23, 2025
    Why does Tuttnauer 1730EK Laboratory Equipment printer feed paper constantly?
    • A
      Alexander MartinSep 23, 2025
      If the printer on your Tuttnauer Laboratory Equipment gives paper feeds all the time when the machine is switched on, make sure the ‘feed button’ on the printer is not stuck.
  • N
    Nicole PetersonSep 7, 2025
    What to do if the keyboard of my Tuttnauer 1730EK Laboratory Equipment does not respond?
    • J
      jennifer13Sep 7, 2025
      If the keyboard of your Tuttnauer Laboratory Equipment does not respond, press the STOP key for 2 seconds to abort any program that may still be running.
  • F
    Frank GonzalezSep 8, 2025
    How to troubleshoot a Tuttnauer 1730EK Laboratory Equipment system that shows no response when turned on?
    • K
      kevinwalkerSep 8, 2025
      If your Tuttnauer Laboratory Equipment does not respond when you turn it on, try the following: First, check and fix all connections. If the issue persists, replace any faulty cables, connectors, or the power switch.
  • K
    Kimberly Peterson DVMSep 23, 2025
    Why is my Tuttnauer Laboratory Equipment making a loud noise?
    • A
      adamrosarioSep 23, 2025
      If you hear a loud noise coming from the reservoir of your Tuttnauer Laboratory Equipment, the sterilization temperature may be set too high. The maximum setting is 274ºF (134ºC).
  • K
    Kathryn HayesAug 31, 2025
    How to clean a very dirty Tuttnauer 1730EK Laboratory Equipment chamber?
    • A
      arojasAug 31, 2025
      To clean a very dirty chamber in your Tuttnauer Laboratory Equipment, use Chamber Brite ™. Very dirty machines may require two cleanings.