Specifications
A.General Specifications:
l Maximum voltage including transient overvoltage
between any terminals and grounding: 500V rms.
l Display: 3 1/2 digits LCD display, Maximum display 1999
l Auto Polarity Display
l Overloading : Display OL or –OL
l Battery Deficiency : Display
l Measurement Speed : Updates 3 times/second.
l Measuremnet Deviation : When the conductor being
meaured is not placed in a
correct position during AC
current measurement, it will
cause
1% reading deviation.
l Drop Test : 1 meter drop test passed.
l Max. Jaw Size : 28mm diameter.
l Projected Max.
Current conductor size : 26mm diameter.
l Power : 2pcs of 1.5V battery (AAA)
l Battery Life : typically 150hours (alkaline
battery)
l Sleep Mode (can be disabled)
l Dimensions (H x W x L) : 30mm x 76mm x 208mm.
Weight : Approximate 260g (battery
included)
B.Environmental Restrictions:
l The Meter is suitable for indoor use.
l Altitude : Operating: 2000m
Storage: 10000m
l Safety/ Compliances : IEC 61010 CAT.II 600V,
CAT.III 300V over voltage
and double insulation standard.
l Temperature and humidity:
Operating: 0
o
C~30
o
C ( 75%
R.H);30
o
C~40
o
C ( 70%R.H);
40
o
C~50
o
C ( 45%R.H);
Storage: -20
o
C~+60
o
C
( 75%R.H)