B. AC Voltage
Range
6.600V
66.00V
600.0V
Resolution
1mV
10mV
100mV
Accuracy
(1.2%+5)
Overload protection
600V DC/AC
Remarks:
Input Impedance: 10M
Frequency Response: 40Hz~400Hz
Change to AC:
Combine AC and True RMS response method. Input sine wave to adjust. Non
sine wave must follow the below data to adjust:
Peak factor: 1.4~2.0, add 1.0% on the stated accuracy
Peak factor: 2.0~2.5, add 2.5% on the stated accuracy
Peak factor: 2.5~3.0, add 4.0% on the stated accuracy.
Model UT207/208: OPERATING MANUAL
41