6. TROUBLE SHOOTING
Syndrome Scale cannot be turned on
Check: Is the scale charged?
Action: Recharge the scale for at least 8
hours before first time use or plug
in the power adaptor before power
on the scale.
Check Is the power adaptor inserted
properly into both the electricity
outlet and the DC inlet of scale.
Action: Secure both ends of the power adaptor
and try again.
Syndrome Scale turned off automatically
Check Is the Auto Power Off function
employed?
Action Refer to 4.2.2 to disable the Auto
Power Off function.
Check Is the LO-BAT indicator on?
Action Apply the power adaptor.
Syndrome Rated capacity cannot be reached
Check Is the TARE indicator on?
Action: Turn the scale off. Remove all loads
from platform and turn on again.
Check Is there anything obstructing the
platform or pillar?
Action Remove all obstacles.
Syndrome Scale displays
00000
Check Does the load applied to platform
excess the rated capacity of scale?
Action Remove all loads from platform and
try again.
Syndrome Weighing result is not accurate
Check Is the scale placed in a level
condition?
Action Adjust the adjustable feet to a level
condition.
Check Is the scale affected by airflow,
vibration or RFI?
Action: Place the scale away from all
disturbances.
Check Is the scale calibrated correctly?