12
Notes concerning repair
Separatefrom
mains
Testing
BVCcontrol/G
BVCprofessional
/G
Genuinespare
parts
3.2 Prior to maintenance and repair
Preparation
> Useonlygenuinesparepartsandaccessories.
Whenusingcomponentsofothermanufacturersthesafetyand
performance of the equipment as well as the electromagnetic
compatibilityoftheequipmentmightbereduced.
PossiblytheCEmarkortheC/USconformity(seeratingplate)
becomesvoidifnotusinggenuinespareparts.
> Checkthattherequiredtoolsandpartsareavailableandof
thecorrecttypebeforeyoustartyourwork.
> Checktheoperatingsequencementallyusingexplodedview
drawings,sparepartslistsandcircuitdiagramonfeasibility,
safetyrequirementsandconsequencesonsafetyandfunc‑
tionoftheequipment.
> Ventthedevice.
> Allowthedevicetocooldown.
> Separatethedevicefrommainspriortomaintenanceor
repair.Switchothedeviceandunplug.
> Allowthecapacitorstodischarge.
> Connectthepersoninchargetogroundpriortoanycontact
withelectroniccomponentstoavoiddamagefromelectro‑
staticdischarges(ESD).
3.3 After maintenance and repair
Testing
Checkoperabilityandsafetyaftermaintenanceandrepair.
> ChecktheelectricalsafetyaccordingtoIEC61010and
nationalregulationsaftermaintenanceandrepair.
BVC control / G und BVC professional / G
> Checktheultimatevacuumofthepumpaftermaintenance
andrepair.