Emptying the collection bottle of the BVC control and
BVC control G
Check liquid level in the collection bottle regu-
larly.
Maximum admissible liquid level in collec-
tion bottle: approx. 80 %, depending on the
application. (In case of low boiling liquids or in
case of suction of liquids which tend to foam,
the admissible liquid level might be reduced.)
max.
80%
Remove screw cap from the collection bottle af-
ter venting.
Remove bottle from the support.
Sterilize and dispose of collected liquid accord-
ing to appropriate regulations.
Remove the connecting tube at the pump inlet,
isolate connections in case.
Switch off the BVC.
Vent the collection bottle.
atmo-
spheric
pressure