Vaisala HMD20 - Calibration Steps (Quick Guide)

Vaisala HMD20
14 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HMI41 Calibration Option
Appendix 1 19165ZZ-U220en-1.1
2 1996-08-26
2. Calibration
After having connected the cable, turn the HMI41 on and follow the
instructions of the table below:
DISPLAY WHAT TO DO PRESS:
RH
setminmax hyst hold Lo batHi
Pabs Pa
g/m
T
Td °C°F
°C°F
%
The HMI41 has been turned on.
Indication of the software version (if the
version is 1.02 or more)
bat
Hi
Indication of the battery voltage.
RH %
Let the readings stabilize and then adjust
the transmitter potentiometers until the
readings are the same.
offset: <65 %RH
gain : >65 %RH
Alternatively, you can adjust the
transmitter when the display shows the
difference in readings.
MODE to have the difference
in readings on the display
ON/OFF if you have
completed the calibration
RH %
Let the readings stabilize and then adjust
the transmitter potentiometers until the
reading on the first line is as close to zero
as possible.
offset: <65 %RH
gain : >65 %RH
MODE to return the measured
values on the display
ON/OFF if you have
completed the calibration