    Andrew JimenezAug 10, 2025
    Why is the temperature reading shown by my Vaisala Transmitter too high?
    • C
      cameron77Aug 10, 2025
      The temperature reading may be too high because the transmitter is installed in an unsuitable location, such as near a heat source or in sunlight. Use a portable reference meter to check this. For HMW82 and HMW83 models, ensure that the transmitter is installed in the proper orientation, with the arrow on the mounting base pointing up.
  • J
    Jose DavisAug 15, 2025
    What to do if my Vaisala INTERCAP HMDW80 outputs an incorrect humidity reading?
    • K
      Kimberly GatesAug 15, 2025
      If your transmitter outputs an incorrect humidity reading: * For models with probes: check if the yellow transport protection cap has been left over the probe. Remove it if present. * For models with a configurable output parameter: check that the desired output parameter has been selected using the DIP switches.
  • T
    Thomas GarzaAug 19, 2025
    What does ERROR HUM mean on my Vaisala INTERCAP HMDW80 Transmitter?
    • D
      dylan32Aug 19, 2025
      The ERROR HUM message indicates a problem with the humidity sensor. Check if the humidity sensor is damaged or missing.
  • R
    Ruben FreyAug 26, 2025
    What does ERROR T mean on my Vaisala INTERCAP HMDW80 Transmitter?
    • A
      Andrew BartonAug 26, 2025
      The ERROR T message indicates a problem with the temperature sensor. Check if the temperature sensor is damaged or missing.
  • N
    Natalie MartinAug 29, 2025
    Why is the current output on the temperature channel ~24 mA on my Vaisala INTERCAP HMDW80?
    • S
      Sean WatsonAug 30, 2025
      If the current output on the temperature channel is approximately 24 mA, check that the humidity output channel is powered. For loop-powered HMDW80 series models that measure humidity, the humidity output channel must be connected, even when only measuring temperature.
  • V
    Victor BenjaminSep 3, 2025
    What does INTERNAL 1 INTERNAL 2 INTERNAL 3 mean on my Vaisala Transmitter?
    • P
      Patrick JonesSep 3, 2025
      If you see 'INTERNAL 1 INTERNAL 2 INTERNAL 3', it indicates an internal problem with the transmitter. Restart the transmitter to see if the problem remains.

Vaisala INTERCAP HMDW80 Series Specifications

General IconGeneral
Protection ClassIP65
Measurement ParametersHumidity and Temperature
Humidity Measurement Range0 to 100% RH
Power Supply10 ... 35 VDC
Response Time15 s (humidity, 90% step at +25°C and 1 m/s airflow)
MaterialStainless steel
Humidity SensorVaisala INTERCAP®
Output Signal4 to 20 mA