Chapter 6 _____________________________________________________________ Technical Data
VAISALA ________________________________________________________________________ 43
CHAPTER 6
TECHNICAL DATA
This chapter provides the technical data of the HMDW80 series.
Specifications
Table 9 Specifications for Models HMW82/83/HMW82P100
and TMW82/83
Property
Description / Value
Operating environment
Operating temperature range
-5 ...+55 °C (+23 ... +131 °F)
Operating humidity range
0 …100 %RH, non-condensing
Relative humidity
Measurement range
0 … 100 %RH
Accuracy
Temperature range
+10 ... +30 °C (+50 ... +86 °F)
30 ... 70 %RH
±3 %RH
0 … 30 %RH, 70 … 100 %RH
±5 %RH
Temperature range
-5 ... +10 °C, +30 ... + 55 °C
(+23 ... +50 °F, +86 ... +131 °F)
0 …100 %RH
±7 %RH
Stability in typical HVAC applications
±2 %RH over 2 years
Humidity sensor
Vaisala INTERCAP
®
Temperature
Measurement range
-5 ...+55 °C (+23 ... +131 °F)
Accuracy
+10 … +30 °C (+50 °F … +86 °F)
±0.5 °C (±0.9 °F)
-5 … +10 °C, +30 … +55 °C
(+23 ... +50 °F, +86 ... +131°F)
±1.0 °C (±1.8 °F)
Temperature sensor
Digital temperature sensor
HMW82P100: Pt100 Class F 0.1
IEC 60751

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala INTERCAP HMDW80 Series and is the answer not in the manual?

Vaisala INTERCAP HMDW80 Series Specifications

General IconGeneral
BrandVaisala
ModelINTERCAP HMDW80 Series
CategoryTransmitter
LanguageEnglish

Related product manuals