EasyManua.ls Logo

Vaisala HPP272 - Measurement Data Registers

Vaisala HPP272
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Address
(Decimal)
Address
(Hexadecimal)
Register Description Data
Format
Unit
260 0103
hex
Relative humidity 16-bit
signed
integer
%RH *100
261 0104
hex
Absolute hydrogen peroxide 16-bit
signed
integer
mg/m
3
264 0107
hex
Water concentration by volume 16-bit
signed
integer
ppm
265 0108
hex
Water vapor pressure 16-bit
signed
integer
hPa
266 0109
hex
Absolute humidity (water) 16-bit
signed
integer
g/m
3
267 010A
hex
Water vapor saturation pressure 16-bit
signed
integer
hPa
268 010B
hex
Temperature from the temperature
probe
16-bit
signed
integer
°C * 100
A.4.2 Configuration Registers
Default power-up values (registers 773 ... 775) are written into non-
volatile EEPROM memory. The EEPROM memory implementation has a limit of
30000 writes, and is intended to be used only when saving long-term or
permanent configurations. Use the volatile memory (registers 769 ... 771, values
cleared on power-up) for non-permanent configurations (for example, if the
probe is used in a system that regularly updates the given values).
CAUTION!
HPP272 User Guide M211972EN-B
48

Table of Contents

Related product manuals