EasyManuals Logo

Vaisala HPP272 User Manual

Vaisala HPP272
66 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #52 background imageLoading...
Page #52 background image
Address
(Decimal)
Address
(Hexadecimal)
Register Description Data
Format
Unit / Valid Range
779 030A
hex
Measurement filtering
factor (does not aect
temperature, water, or RH
measurement)
32-bit float Range: 0.01 ... 1
1 = Filter is disabled
< 1 = Reading is a
combination of latest
measurement and the
earlier reading. The value
of the register defines the
portion of the latest
measurement, for
example, 0.9 means the
reading consists 90 % of
the latest measurement
and 10 % of the earlier
reading.
781 030C
hex
Low H
2
O
2
clipping limit 32-bit float ppm
Any numerical value
(default: 0)
1283 0502
hex
Purge status / manual
start
16-bit
function
status
When reading from
register:
1 ... 100 = Purge is in
progress (progress
shown as 1 ... 100%)
0 = Previous purge
completed successfully,
purge is not in progress
... -1 = Previous purge has
failed
When writing to register:
1 = Start purge
1285 0504
hex
Pressure compensation
on/o
16-bit
boolean
1 = On
0 = O
HPP272 User Guide M211972EN-B
50

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Vaisala HPP272 and is the answer not in the manual?

Vaisala HPP272 Specifications

General IconGeneral
BrandVaisala
ModelHPP272
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals