EasyManua.ls Logo

Vaisala PTB220 Series - OPEN Setting a Barometer from POLL Mode to STOP Mode

Vaisala PTB220 Series
113 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER 6________________________________________________________________ OPERATING
VAISALA _________________________________________________________________________ 49
OPEN Setting a barometer from POLL mode to STOP mode
CLOSE Setting a barometer from STOP mode to POLL
mode
OPEN a <cr> CLOSE <cr>
where:
a = the address of the barometer (0 ... 99)
These commands are used to set a barometer temporarily to STOP
mode and back to POLL mode again. The OPEN command is useful
when several barometers are connected to one serial bus and
communication with a single barometer is needed using all commands
(except for the ? command, see Diagnostic commands on page 66).
The CLOSE command is a global command and requires no address.
The PTB220 series digital barometers must always be closed after the
POLL mode selection. The RESET command and powering-up will
close the PTB220 series barometers automatically.
Example:
>addr 7 <cr>
Address : 7 <cr>
>smode poll <cr>
Serial mode : POLL
>close <cr>
line closed
Example of opening and closing a barometer with address 7:
open 7 <cr> (text invisible)
PTB 7 line opened for operator commands
>
Any commands (except for the ? command) can now be used for the
barometer with address 7 without affecting other barometers with
different addresses connected to the same serial interface. However, if
the RESET command is given or the barometer is powered-up, the
barometer will automatically return to the POLL mode.

Table of Contents

Related product manuals