USER'S GUIDE_______________________________________________________________________
86 ___________________________________________________________________ M210194EN-A
>pulse <cr>
ON SLOW LOW 0.0
>ptest 1000 <cr> (outputting pre-defined pressure 1000.0 hPa)
1000.000 1000.000 10000
>pulse slow low -800 <cr>
ON SLOW LOW -800.0
>ptest 1000 <cr>
1000.000 200.000 2000
>
NOTE
In case of an error in a PTB220 series barometer, there will be no
pulse output from the barometer despite an external trigger pulse.
This error handling technique ensures that a host system can not
receive erroneous pressure readings from a PTB220 series
barometer.
NOTE
If the analog output module is installed the PULSE output mode
cannot be used.