uvSVX and uvSVXe Operating Manual: MANUAL-857569191-11 | issue 1.7
© 2021 Valeport Ltd
6 Data Output Formats
The uvSVX has a selection of different data output formats allowing easy interface to software
packages and third-party instrumentation.
6.1 Data String Formatting
Code
Description
#026;n
Set the string delimiter
Where n = "ASCII character"
e.g. #026;,
#026;/
this function is password protected
#027
Read the data string delimiter character
#042;n
Enable | Disable a leading delimiter character at the beginning of a data string
#042;0 disables the character from the output data strings
#042;1 enables the character in the output data strings
this function is password protected
#043
Read leading data string delimiter character activation status
#044
Enable | Disable a trailing delimiter character at the end of a data string
#044;0 disables the character from the output data strings
#044;1 enables the character in the output data strings
this function is password protected
#045
Read leading data string delimiter character activation status
#083;n
Set the number of decimal places for Pressure / Depth output data string
where n = 1 to 3
this function is password protected
#084
Read the number of decimal places in Pressure / Depth output data string
#116;n
Set the Pressure / Depth error message displayed in the data string for faulty or no pressure
transducer
0 = 0.000
1 = -99999
2 = ERROR
#117
Read the error message