PaX-i PACKING GUIDE
STEP 1. Prepare the Device for Packing Remark Check
In Capture software, press Ctrl + S to bring up Serial
Communication window
Please mark
X’ in box
after you
complete each
step.
Enter ‘pken_0001’ in upper right box and Click ‘SEND’ to
put the unit in Packing Mode
o Once you hear the chime and see [epm_pken_001], the
device is moving into packing mode
o Restart the unit to confirm the unit does not initalize
A Unit moves into
locking position once
you click ‘SEND’
Install the 2 M8x45 locking bolts as shown below to secure
the rotator from moving
Locking Bolts are
included in the Empty
Box
This is to prevent
the Rotator from
moving during
transportation.
If not locked, the
device will be
mechanically
damaged.