PaX-i3D Green (PHT-60CFO) User Manual
152
Appendices
2.2
X-ray Leakage Dose
Test Equipment
Instrument Manufacturer Model S/N
Radiation Monitor
Controller
Radcal 9015 91-1470/19069
Test Condition
Test mode Detector kVP mA sec
PANO/ Adult Xmaru 1501CF 80 16 13.5
CEPH/ 12x10(in.) 1210SGA 99 14 0.7
CT/ 16x10(cm) Xmaru 1524CF Master Plus 99 14 9.0
CT/ 15x15(cm) Xmaru 1524CF Master Plus 99 14 9.0
PANO /
Adult
CEPH /
12x10(in.)
CT /
16x10(cm)
CT /
15x15(cm)
[mR/hr]
0 1.1 5.1 5.2 4.8
45 3.5 14.2 14.6 4.3
90 5.8 18.2 17.0 12.0
100 7.3 14.3 16.0 7.8
110 4.7 15.2 11.7 8.0
120 3.9 22.4 20.3 12.3
130 12.7 51.2 45.1 35.1
140 20.7 45.1 48.7 36.8
150 24.5 57.4 58.1 17.4
160 22.4 51.1 52.8 41.0
170 15.1 33.7 33.5 16.4
180 10.5 33 23.8 16.0
190 5.9 21.4 18.2 5.5
200 3.2 12.9 14.2 14.0
210 2.8 4.3 9.0 3.3
220 3.2 2.7 7.4 5.6
230 3.6 5.5 8.5 2.6
240 4.2 8.2 8.3 4.1
250 6.8 29.7 23.2 21.7
260 4.9 33.3 33.3 18.7
270 19.6 42.1 50.7 25.7
315 3.4 9.4 12.4 11.2
Mode
Direction [°]
[PI3DG_130U_44A_en]User Guide.indd 152 2016-05-24 오후 4:19:23