Page #154 background image
PaX-i3D Green (PHT-60CFO) User Manual
154
Appendices
CT Mode (Xmaru1524CF Master Plus)
Test Condition
Tested Mode
CT FOV 160 x 100
High Resolution Mode 9.0s
Applied Tube Voltage Peak [kVp]
99
Applied Tube Current [mA]
14
FOV 160 x 100 9.0 s [mR]
1 m
(3.3 ft)
1.5 m
(4.9 ft)
2 m
(6.6 ft)
0
1.727 0.971 0.482
45
1.817 0.952 0.499
90
0.656 0.170 0.044
135
2.392 1.121 0.525
180
2.541 1.104 0.480
225
2.399 1.095 0.500
270
2.415 1.185 0.581
315
1.874 0.936 0.467
Mode
Direction [°]