User Manual
Page 15 of 27
[C03 ] Press
[C3 1] d=1
[C3 2] d=2
[C3 5] d=5
[C3 10] d=10
[C3 20] d=20
[C3 50] d=50
Press
, go to next step
C04 Maximum capacity
For example:max weighing 100kg:
set [0100.00]
Press
, go to next step
C05 Zero calibration
option0=non-calibration zero 1=need calibration zero
calibration zero please choose 1 and ensure scale is empty and “stable” light
is on. countdown[CAL 10]~[CAL 0], then the indicator will
show[0.00](example for two decimal point).
C06 Loading calibration
[C06], Press
, show [C06 0], press , change to [C06 1],
Press
, show [SPAn ],
Basic on max capacity setting, add suitable weight on scale. close to the max
capacity, heavier than 10% max at least.
For example: the weight is 80kg