Cleaning
Use non chlorine-based cleaners. Cleaners containing chlorine can cause staining and pitting of the stainless
steel.
Disconnect power to unit by turning the power switch off, located on the rear, right hand corner, and removing the
power cord from the receptacle.
The appliance is designed for medications, vaccines and laboratory product storage so it is important to keep
it clean. The equipment is thoroughly cleaned at the factory before being shipped. We recommend, however, to
clean the interior cabinet before the rst start up of the appliance.
Interior Cleaning
Using a sponge or soft cloth, clean unit with a non-abrasive, non-chlorinated, all-purpose detergent. Note: The
air plenum can be removed to clean it, as well as behind it. In addition, the shelves and slides should be cleaned
aswell.
Exterior Cleaning
Wipe stainless steel exterior with a soft cloth in the direction of grain as needed. Stainless steel polish may be
used to enhance the nish of the unit. The glass door and exterior parts may be cleaned with a soft cloth, window
cleaner or other non-abrasive cleaner.
Annual Cleaning
Removal of dust and other particulates from air intake areas and the condenser is important for proper operation.
Environments with large amounts of dust may require more frequent cleaning.
1. Disconnect power to unit by turning power switch off and removing power cord from receptacle.
2. The condenser can be accessed from rear for cleaning.
3. Inspect drain pan for any debris or obstruction in condensate pan.
4. Use a vacuum cleaner with brush attachment to clean condenser, compressor and related parts.
Refrigeration System
The Vector Lab Products refrigeration system is designed to give many years of trouble-free service. Except for
routine cleaning of the air-cooled condenser and related parts, the refrigeration system requires no service or
maintenance. Refrigerator utilize a capillary tube metering device while freezers utilize a thermal expansion valve.
Both systems are critically charged. Access ttings are not provided. Temporary piercing valves are required to work
on the refrigeration system. Due to the small charge size contained within these products, Vector Lab Products
does not recommend accessing the refrigeration system. If a sealed system failure is suspected, Vector Lab
Products recommends replacing the cooling system in its entirety.