4 Vector Lab Products Refrigerators and Freezers
Ambients
Intended for indoor use where ambient temperature does not exceed 104 F (40C) and ambient relative humidity
does not exceed 65% RH.
Clearances
The top of the unit must have 15 inches of clear space above the refrigerator to ensure proper ventilation of the
refrigeration system.
The back of the refrigerator should have 2 inches of clearance to allow for power cord clearance and ventilation on
unit.
Electrical Specications
§ 115 V, 60 Hz, 1 phase
§ Minimum circuit ampacity: 15A
§ Maximum size of branch circuit overcurrent device: 15A circuit
§ Vector Lab Products recommends circuit be protected by GFCI breaker
§ Does not require dedicated circuit
Refrigeration Specications
§ Refrigerant R290
§ VREF 12: 2.5 oz
§ VREF25: 3.2 oz
§ VREF30: 3.5 oz
§ VREF49: 3.5 oz
§ VREF72: 3.2 oz (2 cooling units)
§ VFZR25: 3.9 oz
§ VFZR49: 3.5 oz