Advanced functions
LFXG-H Installation and Operation Guide 69
Sensor Test
This mode simulates the sensor output at a number of raw counts you
define. This is before application of:
Temperature compensation
Sensor uniformity gain
Standardize gain
The true sensor output is ignored while the transmitter is in sensor test
mode.
This mode is useful for verifying the electronics and software response to
input counts without having to:
Change the process
Shield the source
Vary the radiation field
While in this mode, after entering a number of counts, it may be useful to
look at the Process Chain tab to view the variables affected by the raw
counts value.
Procedure 4.6: To perform a sensor test
Auxiliary Input Test
This mode simulates the auxiliary input frequency at a user-defined number
of counts. The effect of auxiliary input counts depends on the auxiliary input
mode.
Examples:
1. Select Diagnostics | Test | Sensor Test.
2. Click Enter.
3. Remove the gauge from control. Enter the value of the new counts to
force.
4. Click OK.
The transmitter functions in this mode until it times out (1 hour), or
you click Exit and OK.