l Article numbers, documentation
With the serial number, you can access the delivery data of the
instrument via
www.vega.com, "VEGA Tools" and "serial number
search". In addition to the type label outside, you can also nd the
serial number on the inside of the instrument.
3.2 Principle of operation
VEGAFLEX 61 is a level sensor with cable or rod probe for continuous
level measurement.
It is designed for industrial use in all areas of process technology and
can be used equally well in liquids or solids.
High frequency microwave pulses are guided along a steel rope or a
rod. Upon reaching the product surface, the microwave pulses are
reected. The running time is evaluated by the instrument and
outputted as distance.
Power supply via the Probus DP/PA segment coupler or VEGALOG
571 EP cards. A two-wire cable according to Probus specication
serves as carrier of both power and digital data transmission for
multiple sensors. The instrument prole of VEGAFLEX 61 corresponds
to prole specication version 3.0.
The backlight of the indicating and adjustment module is powered by
the sensor. Prerequisite is a certain level of operating voltage.
The data for power supply are spec ied in chapter "Technical data".
The optional heating requires its own operating voltage. You can nd
details in the supplementary instructions manual "Heating for indicat-
ing and adjustment module".
This function is generally not available for approved instruments.
The GSD (instrument master les) and bitmap les necessary for
planning your Probus-DP-(PA) communication network are available
from the download section on the VEGA homepage
www.vega.com
under "Services - Downloads - Software - Probus". There you can
also nd the appropriate certicates. In a PDM environment, an EDD
(Electronic Device Description) is also required to enable the full range
of sensor functions (also available as a download).A CD with the
appropriate les can be ordered via e-mail under info@de.vega.com or
by phone from one of the VEGA agencies under the order number
"DRIVER.S".
3.3 Operation
VEGAFLEX 61 can be adjusted with dierent adjustment media:
l with indicating and adjustment module
Application area
Functional principle
Power supply and bus
communication
GSD/EDD
8 VEGAFLEX 61 Probus PA