9
3 Mounting
Polarisation lock VEGAPULS 64, 69 • 4 … 20 mA
54812-01-191211
4
3
1 2
Abb. 10: Housing lock with Aluminium housing
1 Single chamber housing
2 Double chamber housing
3 Housing
4 Housing lock
1 2
4
3
Abb. 11: Housing lock with precision-cast stainless
steel housing
1 Single chamber housing
2 Double chamber housing
3 Housing
4 Housing lock
2
1
Abb. 12: Housing lock with electropolished stainless
steel housing
1 Housing
2 Housing lock
3 Mounting
3.1 Mounting preparations
Tool
Required tools:
•
Hexagon socket wrench (size 4)
3.2 Installation procedure
Proceed as follows:
1. Loosen the hexagon screws and dismantle
the clamp collars
2. For sensors with plastic horn antenna and
40 mm clamping diameter, use a adjusting
ring.
3. Assemble the clamp collars on the housing
- process tting junction
2
1
4
3
Abb. 13: Mounting the housing lock
1 Housing
2 Clamp collars
3 Adjusting ring
4 Screws
4. Tighten hexagon screws evenly in two or
three passes. Tightening torque, see chap-
ter "Technical data".
5. Check if the housing is securely tightened
against the process tting