The selection "Standpipe" opens a new window in which the inner
diameter of the applied standpipe is entered.
Following the characteristics of the applications and the metrological
features of the sensor are described.
Note:
Probably the operation of the instrument in these applications is sub-
ject to national restrictions in respect to the radio license (see chapter
"For your safety"):
Plastic tank
Transportable plastic tank
Open water (gauge measurement)
Openume(owmeasurement)
Rain water overfall (weir)
Storage tank:
Setup: large-volumed, upright cylindrical, spherical
Mediumspeed:slowllingandemptying
Process/measurement conditions:
Condensation
Smooth product surface
High requirements on measurement accuracy
Properties, sensor:
Low sensitivity to sporadic false echoes
Stable and reliable measured values through averaging
High accuracy
Short reaction time of the sensor not required
Storage tank with product circulation:
Setup: large-volumed, upright cylindrical, spherical
Mediumspeed:slowllingandemptying
Installations: small, laterally mounted or large, top mounted stirrer
Process/measurement conditions:
Relatively smooth product surface
High requirements on measurement accuracy
Condensation
Slight foam generation
Overllingpossible
Properties, sensor:
Low sensitivity to sporadic false echoes
Stable and reliable measured values through averaging
High accuracy, because not set for max. speed
False signal suppression recommended