Vernacare Vortex - Fault 4: Unable to Close the Lid; Fault 5: Unable to Open the Lid

Vernacare Vortex
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
34
Unable to
close the lid
Interlock faulty
Check interlock operates when lid
closed - Replace
Check sensor magnet in handle is
present & clean.
Check hall effect sensor -Replace.
Check foot sensor- Replace.
Arm alignment not
correct.
Check hinge block fasteners are secure
– Tighten.
Check hinge arm bushes for wear –
Replace.
Check arm fasteners are secure –
Tighten.
Check the lid seal for compression when
lid is closed – Replace seal.
Check seal face is free from debris –
Clean.
Unable to
open the lid
Interlock pin not
disengaging
Interlock solenoid jammed – Replace.
Check interlock pin movement is free –
Adjust.
Check foot sensors function – replace.
Check output from PCB to solenoid –
Replace PCB.
Hydraulic damper
failed.
Remove hydraulic damper - Replace
Vent tubing blocked
Clear vent pipe work.
Non-return valve is not
operating correctly.
Check operation of non-return valve in
vent connector for sealing – Replace.
Check service drain - slow drains can
cause a vacuum in the hopper.
Fault 4
Fault 5