EasyManua.ls Logo

Vernacare Vortex - Fault 11: Deodoriser Not Working; Fault 12: Display Screen Blank

Vernacare Vortex
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
38
No deodoriser
being
dispensed.
Blockage in feed
pipework.
Check pipework for blockage- clean.
Dispensing valve
failure.
Check solenoid valve for blockage
clean.
Check non return valve for blockage
clean.
Deoderisor Tank empty
Check fill level and display icon
functions.
Check level switch - replace.
Check non return valve for blockage
clean.
Display screen
blank.
No power to the
screen.
Check connections from pcb and at the
display ensure all pins engaged -
replace
Check power is on at the pcb-
illuminated neon.
Display blank replace display.